Anal Bioanal Chem ; 406(29): 7473-82, 2014 Nov.
Article in English | MEDLINE | ID: mdl-25335819

ABSTRACT

The performance of glow discharge optical emission spectroscopy and mass spectrometry for oxygen determination is investigated using a set of new conductive samples containing oxygen in the percent range in three different matrices (Al, Mg, and Cu) prepared by a sintering process. The sputtering rate corrected calibrations obtained at standard conditions for the 4 mm anode (700 V, 20 mA) in GD-OES are matrix independent for Mg and Al but not for Cu. The importance of a "blue shifted" line of oxygen at 130.22 nm (first reported by Köster) for quantitative analyses by GD-OES is confirmed. Matrix-specific calibrations for oxygen in GD-MS are presented. Two source concepts-fast flow (ELEMENT GD) and low gas flow (VG9000)-are evaluated obtaining higher sensitivity with the static flow source. Additional experiments using Ar-He mixtures or µs pulsed GD are carried out in ELEMENT GD aiming to improve the oxygen sensitivity.