Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
Catheter Cardiovasc Interv ; 95(4): 783-790, 2020 03 01.
Article in English | MEDLINE | ID: mdl-31062927

ABSTRACT

OBJECTIVES: To compare the incidence of permanent pacemaker implantation (PPI) with the CoreValve and Evolut R prostheses, to evaluate the implantation depth with both types of prostheses, and to study factors predicting the need for PPI. BACKGROUND: The Evolut R CoreValve can be recaptured and repositioned during deployment, allowing a more precise implantation. METHODS: A total of 208 patients treated with CoreValve and 137 patients treated with Evolut R were analyzed. The depth of the prosthesis in the LVOT was measured by angiography in the annular perpendicular view projection after deploymen in all patients. RESULTS: Baseline conduction abnormalities were comparable between the groups (85/208, 40.9% vs. 53/137, 38.7%; p = 0.69). The mean prosthesis depth was 10.3 ± 8.6 mm in the CoreValve group and 5.5 ± 2.7 mm in the Evolut R group; p < 0.0001. Conduction disturbances after valve implantation were more frequent with the CoreValve (new-onset left bundle branch block: 93, 44.7% vs. 16, 11.7%; p < 0.05, first-degree atrioventricular block: 23, 11.1% vs. 5, 3.6%; p < 0.05). In addition, the incidence of PPI was significantly lower with Evolut R (45, 21.6% vs. 15, 10.9%; p = 0.01). The predictors of the need for PPI were the mean depth of the prosthesis (OR: 1.13, 95% CI: 1.06-1.21; p < 0.0001) and prior right bundle branch block (OR 10.22, 95% CI: 4.62-22.63; p < 0.0001). CONCLUSIONS: The recapturable capability of the Evolut R system allowed for higher and precise valve implantation. This fact had an impact on the reduction in the need for PPI.


Subject(s)
Aortic Valve Stenosis/surgery , Aortic Valve/surgery , Arrhythmias, Cardiac/therapy , Cardiac Pacing, Artificial , Heart Valve Prosthesis , Pacemaker, Artificial , Transcatheter Aortic Valve Replacement/instrumentation , Aged , Aged, 80 and over , Aortic Valve/diagnostic imaging , Aortic Valve/physiopathology , Aortic Valve Stenosis/diagnostic imaging , Aortic Valve Stenosis/physiopathology , Arrhythmias, Cardiac/diagnosis , Arrhythmias, Cardiac/etiology , Arrhythmias, Cardiac/physiopathology , Female , Humans , Male , Prosthesis Design , Retrospective Studies , Risk Factors , Severity of Illness Index , Time Factors , Transcatheter Aortic Valve Replacement/adverse effects , Treatment Outcome
SELECTION OF CITATIONS
SEARCH DETAIL