Radiology
; 178(2): 393-6, 1991 Feb.
Article
in English
| MEDLINE
| ID: mdl-1987599
ABSTRACT
An automatic exposure controller has been designed that controls the optical film density for film, screen, and radiographic techniques typically used in mammography to within 0.05 over a range of 1.3-6.7-cm thickness of Lucite. This degree of accuracy is better than that reported for presently available controllers. The detector system consists of four cadmium telluride detectors and involves the use of a control algorithm to read the detectors and turn off the mammography unit at the correct time. This algorithm is implemented by a microprocessor, which also provides the means for a convenient calibration.
Subject(s)
Mammography/instrumentation , Technology, Radiologic , Female , Humans , Mammography/methods