1.
J Arthroplasty ; 20(7 Suppl 3): 63-72, 2005 Oct.
Article in English | MEDLINE | ID: mdl-16214005

ABSTRACT

Using 743 total hip arthroplasties that included 6 hemispheric porous-coated cup designs, this study used a multiple linear regression to identify those factors that influenced polyethylene wear rates. Wear rates for each hip were based on serial head penetration measurement made with computer-assisted techniques. Implant factors associated with an increased wear rate included terminal sterilization with a non-cross-linking chemical surface treatment, a 4-mm lateralized liner, a cobalt-chrome femoral head, and a longer shelf life for liners gamma-irradiated in air. After accounting for these implant characteristics and patient factors, wear rates among the 6 cup designs were not significantly different (P = .89). Although polyethylene wear is frequently characterized for specific implant designs, this study demonstrated that there are several common factors that influence polyethylene wear rates.


Subject(s)
Hip Prosthesis , Polyethylene , Acetabulum , Adult , Aged , Aged, 80 and over , Arthroplasty, Replacement, Hip , Female , Humans , Male , Materials Testing , Middle Aged , Prosthesis Design