Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Language
Publication year range
1.
Article in English | WPRIM (Western Pacific) | ID: wpr-891063

ABSTRACT

Purpose@#We sought to describe and analyze the adverse events associated with synthetic male slings reported to the U.S. Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database. @*Methods@#We queried the MAUDE database for all entries including the terms “Male Sling,” “InVance,” “Virtue,” or “Advance” from January 1st, 2009 to December 31st, 2018. We collected and analyzed information about the event type, date received, report source, source type, and manufacturer. We reviewed and categorized the event description text for each medical device report (MDR). @*Results@#A total of 497 adverse events related to the male sling were identified. The adverse events were classified as injury (95.4%), malfunction (4.2%), and other (0.4%). There were no deaths described. The slings involved were the Advance or Advance XP sling (69.8%), InVance (15.5%), Virtue Quadratic (12.3%), or unknown (2.4%). The 4 most common adverse events described were urinary incontinence (46.7%), sling erosion (9.1%), mechanical malfunction (8.2%), and painumbness (8.2%). There was no increase in the number of reports in the years following the FDA warnings for urogynecologic mesh. @*Conclusions@#There was an overall modest number of MDRs related to male slings and the majority of them were reported by the manufacturer. The reporting of adverse events for male slings does not seem to be affected by the controversy and scrutiny towards transvaginal mesh and midurethral slings. Further clinical studies and more objective and detailed databases are needed to investigate the safety of these synthetic slings.

2.
Article in English | WPRIM (Western Pacific) | ID: wpr-898767

ABSTRACT

Purpose@#We sought to describe and analyze the adverse events associated with synthetic male slings reported to the U.S. Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database. @*Methods@#We queried the MAUDE database for all entries including the terms “Male Sling,” “InVance,” “Virtue,” or “Advance” from January 1st, 2009 to December 31st, 2018. We collected and analyzed information about the event type, date received, report source, source type, and manufacturer. We reviewed and categorized the event description text for each medical device report (MDR). @*Results@#A total of 497 adverse events related to the male sling were identified. The adverse events were classified as injury (95.4%), malfunction (4.2%), and other (0.4%). There were no deaths described. The slings involved were the Advance or Advance XP sling (69.8%), InVance (15.5%), Virtue Quadratic (12.3%), or unknown (2.4%). The 4 most common adverse events described were urinary incontinence (46.7%), sling erosion (9.1%), mechanical malfunction (8.2%), and painumbness (8.2%). There was no increase in the number of reports in the years following the FDA warnings for urogynecologic mesh. @*Conclusions@#There was an overall modest number of MDRs related to male slings and the majority of them were reported by the manufacturer. The reporting of adverse events for male slings does not seem to be affected by the controversy and scrutiny towards transvaginal mesh and midurethral slings. Further clinical studies and more objective and detailed databases are needed to investigate the safety of these synthetic slings.

SELECTION OF CITATIONS
SEARCH DETAIL
...