Technological advances have made it possible to seamlessly integrate modern neuroimaging into the neurosurgical operative environment. This integration has introduced many new applications improving surgical treatments. One major addition to the neurosurgical armamentarium is intraoperative navigation and MRI, enabling real-time use during surgery. In the 1970s, the American College of Radiology issued safety guidelines for diagnostic MRI facilities. Until now, however, no such guidelines existed for the MRI-integrated operating room, which is a high-risk zone requiring standardized protocols to ensure the safety of both the patient and the operating room staff. The forces associated with the strong 1.5- and 3.0-T magnets used for MRI are potent and hazardous, creating distinct concerns regarding safety, infection control, and image interpretation. Authors of this paper provide an overview of the intraoperative MRI operating room, safety considerations, and a series of checklists and protocols for maintaining safety in this zero tolerance environment.


Subject(s)
Checklist , Magnetic Resonance Imaging/methods , Neurosurgical Procedures/methods , Anesthesia , Computer Systems , Electromagnetic Fields , Humans , Magnetic Resonance Imaging/adverse effects , Magnetic Resonance Imaging/instrumentation , Medical Errors/prevention & control , Operating Rooms/organization & administration , Patient Safety