JACC Heart Fail ; 3(10): 818-28, 2015 Oct.
Article in English | MEDLINE | ID: mdl-26450000

ABSTRACT

OBJECTIVES: The purpose of this study was to determine the risk factors for ischemic in hemorrhage cerebrovascular events in patients supported by the HeartWare ventricular assist device (HVAD). BACKGROUND: Patients supported with left ventricular assist devices are at risk for both ischemic and hemorrhagic cerebrovascular events. METHODS: Patients undergoing implantation with a HVAD as part of the bridge-to-transplant trial and subsequent continued access protocol were included. Neurological events (ischemic cerebrovascular accidents [ICVAs] and hemorrhagic cerebrovascular accidents [HCVAs]) were assessed, and the risk factors for these events were evaluated in a multivariable model. RESULTS: A total of 382 patients were included: 140 bridge-to-transplant patients from the ADVANCE (Evaluation of the HeartWare Left Ventricular Assist Device for the Treatment of Advanced Heart Failure) clinical trial and 242 patients from the continued access protocol. Patients had a mean age of 53.2 years; 71.2% were male, and 68.1% were white. Thirty-eight percent had ischemic heart disease, and the mean duration of support was 422.7 days. The overall prevalence of ICVA was 6.8% (26 of 382); for HCVA, it was 8.4% (32 of 382). Pump design modifications and a protocol-driven change in the antiplatelet therapy reduced the prevalence of ICVA from 6.3% (17 of 272) to 2.7% (3 of 110; p = 0.21) but had a negligible effect on the prevalence of HVCA (8.8% [24 of 272] vs. 6.4% [7 of 110]; p = 0.69). Multivariable predictors of ICVA were aspirin ≤81 mg and atrial fibrillation; predictors of HCVA were mean arterial pressure >90 mm Hg, aspirin ≤81 mg, and an international normalized ratio >3.0. Eight of the 30 participating sites had established improved blood pressure management (IBPM) protocols. Although the prevalence of ICVA for those with and without IBPM protocols was similar (5.3% [6 of 114] vs. 5.2% [14 of 268]; p = 0.99), those with IBPM protocols had a significantly lower prevalence of HCVA (1.8% [2 of 114] vs. 10.8% [29 of 268]; p = 0.0078). CONCLUSIONS: Anticoagulation, antiplatelet therapy, and blood pressure management affected the prevalence of cerebrovascular events after implantation of the HVAD. Attention to these clinical parameters can have a substantial impact on the occurrence of serious neurological events. (Evaluation of the HeartWare Left Ventricular Assist Device for the Treatment of Advanced Heart Failure [ADVANCE]; NCT00751972).


Subject(s)
Cause of Death , Heart Failure/surgery , Heart-Assist Devices/adverse effects , Stroke/etiology , Waiting Lists , Aged , Brain Ischemia/etiology , Brain Ischemia/mortality , Brain Ischemia/physiopathology , Female , Follow-Up Studies , Heart Failure/diagnosis , Heart Failure/mortality , Heart Transplantation/methods , Humans , Kaplan-Meier Estimate , Male , Middle Aged , Prospective Studies , Risk Assessment , Severity of Illness Index , Stroke/mortality , Stroke/physiopathology , Survival Rate , Treatment Outcome
2.
JACC Heart Fail ; 3(4): 303-13, 2015 Apr.
Article in English | MEDLINE | ID: mdl-25770405

ABSTRACT

OBJECTIVES: This study evaluated gastrointestinal bleeding (GIB) in patients receiving the HeartWare HVAD System (HeartWare Inc., Framingham, Massachusetts) in the pivotal BTT (Bridge to Transplant) trial and under the continued access protocol (CAP). BACKGROUND: GIB has become a significant problem for recipients of continuous flow device left ventricular assist devices (CF-LVAD). The need for anticoagulation and antiplatelet therapies complicates the management of GIB. METHODS: Bleeding events from 382 patients with advanced heart failure (140 patients enrolled in the BTT trial, and an additional 242 CAP patients) were analyzed. Post-implant anticoagulation consisted of heparin followed by warfarin at a target international normalized ratio of 2 to 3. Acetylsalicylic acid was recommended at 81 to 325 mg. RESULTS: Overall, 59 of 382 (15.4%) patients experienced 108 GIB events (0.27 events per patient year). Mean time to first bleed was 273.1 days and 86.1% of events occurred beyond 30 days. Freedom from GIB was 84.1% at 1 year. Median international normalized ratio at the time of first bleed was 2.4 ± 1.4. The most common etiology of bleeding identified was arteriovenous malformation and the most common site was the small intestine. Repeat bleeding was infrequent, though GIB patients required more readmissions and developed nondevice infections more frequently. No patients required surgical intervention and no deaths directly related to GIB occurred. CONCLUSIONS: Recipients of the HeartWare Ventricular Assist Device System had an incidence of 0.27 GIB/patient year with a freedom from GIB of 84.1% at 1 year. All patients with GIB events were managed with medical and endoscopic therapies, although 31% of patients experienced a recurrence of GIB. No surgical intervention was required. GIB did not impact survival. (Evaluation of the HeartWare Left Ventricular Assist Device for the Treatment of Advanced Heart Failure [ADVANCE]; NCT00751972).


Subject(s)
Gastrointestinal Hemorrhage/etiology , Heart Failure/therapy , Heart-Assist Devices/adverse effects , Anticoagulants/adverse effects , Female , Gastrointestinal Hemorrhage/mortality , Heart Failure/mortality , Heart Transplantation/statistics & numerical data , Humans , Kaplan-Meier Estimate , Male , Middle Aged , Platelet Aggregation Inhibitors/adverse effects , Prospective Studies
3.
J Heart Lung Transplant ; 34(6): 815-24, 2015 Jun.
Article in English | MEDLINE | ID: mdl-25813372

ABSTRACT

BACKGROUND: The Ventricular Assist Device for the Treatment of Advanced Heart Failure (ADVANCE) Bridge to Transplant (BTT) trial was a multicenter, prospective trial of the HeartWare Ventricular Assist Device (HVAD). The performance of the HVAD in various demographic sub-groups was evaluated. METHODS: Baseline characteristics, adverse events, and survival were compared for men vs. women and whites vs. non-whites in the combined ADVANCE BTT and continued access protocol trial. Of 332 patients enrolled in these trials, 236 were men and 96 women, with 228 whites and 104 non-whites. RESULTS: At baseline, women had a smaller body surface area (1.8 ± 0.2 vs. 2.1 ± 0.3 m2, p < 0.0001), less hypertension (50.0% vs. 61.9%, p = 0.05), and less ischemic cardiomyopathy (15.6% vs. 45.3%, p < 0.0001). Differences in Kaplan-Meier survival were not significant at 180 days (men, 91.8%; women, 91.7%) and 1 year (men, 85.3%; women, 85.1%) despite adjustment for baseline differences. Men had a lower incidence of early right heart failure and renal and respiratory dysfunction, and a shorter length of stay. In the analysis by race, non-whites were younger than whites and had less ischemic heart failure, more hypertension, and lower creatinine levels at baseline. Non-whites had lower rates of arrhythmia, bleeding requiring rehospitalization, and device malfunctions than whites. Survival was high in non-whites and whites, at 94.1% vs. 90.4% at 180 days and 89.2% vs. 82.8% at 1 year, respectively, despite adjustment for baseline differences. CONCLUSIONS: Although heart failure etiology differed between men and women and between whites and non-whites, sex and race were not factors that affected survival in patients receiving the HVAD as BTT, which was high in all sub-groups.


Subject(s)
Heart Failure/mortality , Heart Failure/therapy , Heart Transplantation , Heart-Assist Devices , Racial Groups , Sex Factors , Adult , Creatinine/blood , Female , Follow-Up Studies , Heart Failure/ethnology , Humans , Hypertension/epidemiology , Incidence , Kaplan-Meier Estimate , Length of Stay , Male , Middle Aged , Prospective Studies , Retrospective Studies , Survival Rate , Treatment Outcome
4.
Circulation ; 130(11 Suppl 1): S3-11, 2014 Sep 09.
Article in English | MEDLINE | ID: mdl-25200052

ABSTRACT

BACKGROUND: Cardiac valve procedures are commonly performed concurrently during implantation of left ventricular assist devices, but the added procedural risk has not been studied in detail. METHODS AND RESULTS: Data from patients receiving the HeartWare Ventricular Assist Device in the ADVANCE bridge to transplant (BTT) trial and continued access protocol were reviewed. Of 382 consecutive patients who completed follow-up between August 2008 and June 2013 (mean time on support 389 days, median 271 days), 262 (68.6%) underwent isolated HeartWare Ventricular Assist Device implantation, 75 (19.6%) a concurrent valve procedure, and 45 (11.8%) concurrent nonvalvular procedures. Of the concurrent valve procedures, 56 were tricuspid, 13 aortic, and 6 mitral. Survival was similar between groups (79% for concurrent valve procedures and 85% for HeartWare Ventricular Assist Device only at 1 year; P=0.33). Concurrent valve procedures were also associated with increased unadjusted early right heart failure (RHF). A multivariable analysis for death and RHF (121 total events) identified female sex (odds ratio=2.0 [95% confidence interval, 1.2-3.3; P=0.0053]) and preimplant tricuspid regurgitation severity (odds ratio=2.9 [95% confidence interval, 1.8-4.8, P<0.0001]) as independent predictors while concurrent tricuspid valve procedures (TVP) were not predictors. Furthermore, patients with significant preimplant tricuspid regurgitation who did not receive a TVP experienced an increased rate of late RHF compared with those who received TVP (0.19 versus 0.05 events per patient-year, respectively; P=0.024). CONCLUSIONS: Compared with HeartWare Ventricular Assist Device alone, survival was equivalent for the concurrent valve procedure group. Tricuspid regurgitation severity was the most important predictor of increased postoperative RHF, and concurrent TVP was not an independent predictor of RHF overall. Concurrent TVP may reduce the rate of late RHF for patients with significant preimplant tricuspid insufficiency. CLINICAL TRIAL REGISTRATION URL: http://www.clinicaltrials.gov. Unique identifier: NCT00751972.


Subject(s)
Cardiac Surgical Procedures , Clinical Trials as Topic/statistics & numerical data , Heart Failure/prevention & control , Heart Valve Diseases/surgery , Heart-Assist Devices , Multicenter Studies as Topic/statistics & numerical data , Adult , Aged , Female , Follow-Up Studies , Foramen Ovale, Patent/surgery , Heart Failure/etiology , Heart Septal Defects/surgery , Heart Transplantation , Heart Valve Diseases/complications , Heart Valve Prosthesis Implantation , Humans , Male , Middle Aged , Pericardium/surgery , Postoperative Complications/epidemiology , Postoperative Complications/etiology , Retrospective Studies , Treatment Outcome , Tricuspid Valve Insufficiency/complications , Tricuspid Valve Insufficiency/surgery
5.
J Heart Lung Transplant ; 33(10): 1066-73, 2014 Oct.
Article in English | MEDLINE | ID: mdl-25087103

ABSTRACT

BACKGROUND: Drive-line infections and sepsis lead to significant morbidity and even mortality in patients with ventricular assist devices (VADs). The HeartWare HVAD system is unique compared with other VAD systems in that it has a thin, flexible drive-line and is implanted directly into the pericardial space without the need for a pump pocket. We reviewed 332 patients receiving the HVAD in the pivotal ADVANCE Bridge to Transplant (BTT) trial and Continued Access Protocol (CAP) to determine the incidence of drive-line infections and sepsis. METHODS: Summary statistics of outcomes and Kaplan-Meier freedom-from-event analyses were performed. The Cochran-Mantel-Haenszel test for homogeneity was used to determine whether there was any site effect on rates of either drive-line exit-site or sepsis infections. RESULTS: Drive-line exit-site infections occurred in 16.9% (56 of 332) of patients (0.25 event per patient-year [EPPY]). These infections were primarily treated with antibiotics and most occurred >30 days post-implant (88%, or 66 of 75 drive-line infections). Sepsis occurred in 17.2% (57 of 332) of patients (0.23 EPPY), and sepsis events occurred more frequently >30 days post-implant (86%, or 60 of 70 sepsis events). Of patients with sepsis, 17.5% (10 of 57) died due to sepsis-related neurologic events (7 of 10) and multisystem organ failure (2 of 10) and cardiopulmonary failure (1 of 10), whereas there were no deaths related to drive-line infections. In addition, 17.5% (10 of 57) of patients had a stroke event while under treatment for sepsis (7 hemorrhagic and 3 ischemic strokes), and 3.5% (2 of 57 patients with sepsis) had a device exchange for VAD thrombus (4 days before and 1 day after sepsis diagnosis). Patients with drive-line infections or sepsis had a larger body mass index compared to those without infection (29.4 vs 27.6 kg/m(2), p = 0.015), and were more frequently diabetic (43.7% vs 31.2%, p = 0.034). Although there was no negative impact on survival in patients with drive-line exit-site infections, there was a trend for reduced survival in patients with sepsis events when compared to those patients without sepsis. CONCLUSIONS: The incidence of drive-line infection and sepsis was low in patients receiving the HVAD pump. Despite a trend toward reduced survival in patients with sepsis events, the overall survival of patients with either drive-line infections or sepsis was excellent.


Subject(s)
Equipment Contamination , Heart Failure/therapy , Heart-Assist Devices/classification , Heart-Assist Devices/microbiology , Sepsis/epidemiology , Adult , Aged , Anti-Bacterial Agents/therapeutic use , Equipment Design , Female , Follow-Up Studies , Humans , Incidence , Kaplan-Meier Estimate , Male , Middle Aged , Retrospective Studies , Sepsis/drug therapy , Sepsis/mortality , Treatment Outcome
6.
J Heart Lung Transplant ; 33(1): 23-34, 2014 Jan.
Article in English | MEDLINE | ID: mdl-24418731

ABSTRACT

BACKGROUND: The HeartWare left ventricular assist device (HVAD, HeartWare Inc, Framingham, MA) is the first implantable centrifugal continuous-flow pump approved for use as a bridge to transplantation. An infrequent but serious adverse event of LVAD support is thrombus ingestion or formation in the pump. In this study, we analyze the incidence of pump thrombus, evaluate the comparative effectiveness of various treatment strategies, and examine factors pre-disposing to the development of pump thrombus. METHODS: The analysis included 382 patients who underwent implantation of the HVAD as part of the HeartWare Bridge to Transplant (BTT) and subsequent Continued Access Protocol (CAP) trial. Descriptive statistics and group comparisons were generated to analyze baseline characteristics, incidence of pump thrombus, and treatment outcomes. A multivariate analysis was performed to assess significant risk factors for developing pump thrombus. RESULTS: There were 34 pump thrombus events observed in 31 patients (8.1% of the cohort) for a rate of 0.08 events per patient-year. The incidence of pump thrombus did not differ between BTT and CAP. Medical management of pump thrombus was attempted in 30 cases, and was successful in 15 (50%). A total of 16 patients underwent pump exchange, and 2 underwent urgent transplantation. Five patients with a pump thrombus died after medical therapy failed, 4 of whom also underwent a pump exchange. Survival at 1 year in patients with and without a pump thrombus was 69.4% and 85.5%, respectively (p = 0.21). A multivariable analysis revealed that significant risk factors for pump thrombus included a mean arterial pressure > 90 mm Hg, aspirin dose ≤ 81 mg, international normalized ratio ≤ 2, and Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) profile level of ≥ 3 at implant. CONCLUSIONS: Pump thrombus is a clinically important adverse event in patients receiving an HVAD, occurring at a rate of 0.08 events per patient-year. Significant risk factors for pump thrombosis include elevated blood pressure and sub-optimal anti-coagulation and anti-platelet therapies. This suggests that pump thrombus event rates could be reduced through careful adherence to patient management guidelines.


Subject(s)
Heart Failure/therapy , Heart Transplantation , Heart-Assist Devices/adverse effects , Thrombosis/epidemiology , Ventricular Dysfunction, Left/therapy , Adult , Anticoagulants/therapeutic use , Female , Humans , Incidence , Kaplan-Meier Estimate , Male , Middle Aged , Multivariate Analysis , Platelet Aggregation Inhibitors/therapeutic use , Retrospective Studies , Survival Rate , Thrombosis/mortality , Thrombosis/prevention & control , Treatment Outcome
7.
J Heart Lung Transplant ; 32(7): 675-83, 2013 Jul.
Article in English | MEDLINE | ID: mdl-23796152

ABSTRACT

BACKGROUND: The HeartWare Ventricular Assist System (HeartWare Inc, Framingmam, MA) is a miniaturized implantable, centrifugal design, continuous-flow blood pump. The pivotal bridge to transplant and continued access protocols trials have enrolled patients with advanced heart failure in a bridge-to-transplant indication. METHODS: The primary outcome, success, was defined as survival on the originally implanted device, transplant, or explant for ventricular recovery at 180 days. Secondary outcomes included an evaluation of survival, functional and quality of life outcomes, and adverse events. RESULTS: A total of 332 patients in the pivotal bridge to transplant and continued access protocols trial have completed their 180-day primary end-point assessment. Survival in patients receiving the HeartWare pump was 91% at 180 days and 84% at 360 days. Quality of life scores improved significantly, and adverse event rates remain low. CONCLUSIONS: The use of the HeartWare pump as a bridge to transplant continues to demonstrate a high 180-day survival rate despite a low rate of transplant. Adverse event rates are similar or better than those observed in historical bridge-to-transplant trials, despite longer exposure times due to longer survival and lower transplant rates.


Subject(s)
Heart Failure/surgery , Heart-Assist Devices , Clinical Protocols , Equipment Design , Female , Heart Transplantation , Humans , Male , Middle Aged , Prospective Studies , Treatment Outcome , Waiting Lists
8.
Circulation ; 125(25): 3191-200, 2012 Jun 26.
Article in English | MEDLINE | ID: mdl-22619284

ABSTRACT

BACKGROUND: Contemporary ventricular assist device therapy results in a high rate of successful heart transplantation but is associated with bleeding, infections, and other complications. Further reductions in pump size, centrifugal design, and intrapericardial positioning may reduce complications and improve outcomes. METHODS AND RESULTS: We studied a small, intrapericardially positioned, continuous-flow centrifugal pump in patients requiring an implanted ventricular assist device as a bridge to heart transplantation. The course of investigational pump recipients was compared with that of patients implanted contemporaneously with commercially available devices. The primary outcome, success, was defined as survival on the originally implanted device, transplantation, or explantation for ventricular recovery at 180 days and was evaluated for both noninferiority and superiority. Secondary outcomes included a comparison of survival between groups and functional and quality-of-life outcomes and adverse events in the investigational device group. A total of 140 patients received the investigational pump, and 499 patients received a commercially available pump implanted contemporaneously. Success occurred in 90.7% of investigational pump patients and 90.1% of controls, establishing the noninferiority of the investigational pump (P<0.001; 15% noninferiority margin). At 6 months, median 6-minute walk distance improved by 128.5 m, and both disease-specific and global quality-of-life scores improved significantly. CONCLUSIONS: A small, intrapericardially positioned, continuous-flow, centrifugal pump was noninferior to contemporaneously implanted, commercially available ventricular assist devices. Functional capacity and quality of life improved markedly, and the adverse event profile was favorable. CLINICAL TRIAL REGISTRATION: URL: http://www.clinicaltrials.gov. Unique identifier: NCT00751972.


Subject(s)
Heart Transplantation/trends , Heart-Assist Devices/trends , Pericardium , Waiting Lists , Adult , Aged , Female , Follow-Up Studies , Heart Failure/mortality , Heart Failure/physiopathology , Heart Failure/surgery , Heart Transplantation/mortality , Humans , Male , Middle Aged , Pericardium/physiopathology , Prospective Studies , Surveys and Questionnaires , Survival Rate/trends , Treatment Outcome , Ventricular Dysfunction, Left/mortality , Ventricular Dysfunction, Left/physiopathology , Ventricular Dysfunction, Left/surgery , Waiting Lists/mortality
SELECTION OF CITATIONS