Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Magn Reson Imaging ; 20(2): 155-63, 2002 Feb.
Article in English | MEDLINE | ID: mdl-12034336

ABSTRACT

We have mitigated acoustic noise in a 1.5 T cylindrical MRI scanner equipped with epoxy-potted, shielded gradients. It has been widely assumed that MRI acoustic noise comes overwhelmingly from vibrations of the gradient assembly. However, with vibration-isolated gradients contained in an airtight enclosure, we found the primary sources of acoustic noise to be eddy-current-induced vibrations of metal structures such as the cryostat inner bore and the rf body coil. We have elucidated the relative strengths of source-pathways of acoustic noise and assembled a reduced-acoustic-noise demonstration MRI system. This scanner employed a number of acoustic noise reduction measures including a vacuum enclosure of a vibrationally isolated gradient assembly, a low-eddy-current rf coil and a non-conducting inner bore cryostat. The demonstration scanner reduced, by about 20 dBA, the acoustic noise levels in the patient bore to 85 dBA and below for several typical noisy pulse sequences. The noise level standing near the patient bore is 71 dBA and below. We have applied Statistical Energy Analysis to develop a vibroacoustic model of the MR system. Our model includes vibrational sources and acoustic pathways to predict acoustic noise and provides a good spectral match above 400 Hz to experimentally measured sound levels. This tool enables us to factor acoustics into the design parameters of new MRI systems.


Subject(s)
Magnetic Resonance Imaging , Noise , Equipment Design , Humans , Magnetic Resonance Imaging/instrumentation , Vibration
SELECTION OF CITATIONS
SEARCH DETAIL
...