1.
Acta Orthop ; 93: 854-858, 2022 11 24.
Article in English | MEDLINE | ID: mdl-36445034

ABSTRACT

BACKGROUND AND PURPOSE: Long-term outcome of small head (28 mm) metal-on-metal (MoM) total hip arthroplasty (THA) is available mainly for Metasul devices (Sulzer Medica, Winterthur, Switzerland). Biomet MoM THA was frequently used in Finland. Therefore, we assessed long-term survivorship of the M2a 28-mm RingLoc MoM THA (Biomet, Warsaw, IN, USA) and compared it with the metal-on-polyethylene (MoP) RingLoc THA from the same manufacturer. PATIENTS AND METHODS: We conducted a register study based on THAs from the Finnish Arthroplasty Register performed between January 1, 2000 and December 31, 2007. 290 28-mm head M2a MoM THAs and 1,647 28-mm head MoP THAs (reference group) were included. The endpoint was revision for any reason, or revision for aseptic loosening, osteolysis, liner wear, or metallosis as one group. Kaplan-Meier survival estimates were calculated, and revision risks were assessed using a Cox multiple regression model, both with 95% confidence intervals (CI). RESULTS: No difference was found in the 15-year Kaplan-Meier survivorship between the 28-mm head M2a RingLoc MoM THA and the reference group for any reason for revision (87.7% [82.9-92.1] and 83.3% [81.0-85.3], respectively). The adjusted hazard ratio (HR) for any reason for revision for the MoM THA group compared with the reference group was at least equal or better (0.70 [0.48-1.02]). Both groups presented similar survival for revision for aseptic loosening of the cup, osteolysis, liner wear, or metallosis, at 96.2% (92.7-98.0) and 95.4% (93.9-96.5), respectively. INTERPRETATION: In the long-term survival there was no difference between the M2a 28-mm RingLoc MoM THA and 28-mm MoP THA. Further follow-up regimens for M2a 28-mm RingLoc THA patients may be unnecessary, but long-term metal ion and radiological data is needed before any formal suggestions.


Subject(s)
Arthroplasty, Replacement, Hip , Metal-on-Metal Joint Prostheses , Osteolysis , Humans , Polyethylene , Arthroplasty, Replacement, Hip/adverse effects , Finland/epidemiology , Cimetidine , Metal-on-Metal Joint Prostheses/adverse effects , Metals
2.
Acta Orthop ; 92(3): 316-322, 2021 06.
Article in English | MEDLINE | ID: mdl-33517823

ABSTRACT

Background and purpose - The use of crosslinked polyethylene in total hip arthroplasty (THA) has decreased wear remarkably. It has been suggested that the antioxidative effects of vitamin E may enhance the wear properties of polyethylene even further. This study evaluates revision rates between vitamin E-infused polyethylene liners (E1 and E-poly, ZimmerBiomet, Warsaw, IN, USA) versus moderately crosslinked polyethylene (ModXLPE) liners from the same manufacturer used in primary THA.Patients and methods - We conducted a study based on data from the Finnish Arthroplasty Register. The study group consisted of 2,723 THAs with a vitamin E-infused liner and a reference group of 2,707 THAs with a moderately crosslinked polyethylene liner. Survivorship, revision risk, and re-revision causes were compared between groups.Results - The 7-year survival of the vitamin E-infused polyethylene liner group and of the reference group with revision for any reason as the endpoint was comparable (94% [95% CI 92.9-94.9] and 93% [CI 91.9-93.9], respectively). The adjusted hazard ratio (HR) for any revision was similar between the groups (0.7 [CI 0.4-1.1]). When revision for aseptic loosening was studied as the endpoint, the survival for the study group was 99% (CI 98.6-99.4) and for the reference group 99% (CI 98.7-99.5), and the risk of revision was comparable between the study groups (HR 1.3 [CI 0.7-2.5]).Interpretation - After an observation period of 7 years vitamin E-infused liners shows results equal to results obtained with crosslinked polyethylene liners.


Subject(s)
Antioxidants/administration & dosage , Arthroplasty, Replacement, Hip/instrumentation , Hip Prosthesis , Prosthesis Design , Prosthesis Failure , Vitamin E/administration & dosage , Aged , Arthroplasty, Replacement, Hip/adverse effects , Female , Finland , Humans , Kaplan-Meier Estimate , Male , Middle Aged , Osteoarthritis, Hip/diagnosis , Osteoarthritis, Hip/etiology , Osteoarthritis, Hip/surgery , Polyethylene , Proportional Hazards Models , Registries , Reoperation
3.
J Bone Joint Surg Am ; 102(13): 1177-1185, 2020 Jul 01.
Article in English | MEDLINE | ID: mdl-32618925

ABSTRACT

BACKGROUND: To enhance osseointegration in total hip arthroplasty (THA), ultraporous or highly porous-coated cups were introduced. Implant survival data on these new devices have been scarce. The aim of our study was to assess the survivorship of ultraporous Tritanium cups (Stryker) in a population-based register study. METHODS: In this study, we collected data on 6,080 primary THAs using a Tritanium cup and 25,670 THAs using a conventional cup (control group) from the Finnish Arthroplasty Register; these procedures were performed from January 1, 2009, to December 31, 2017. We calculated the Kaplan-Meier survival estimates with 95% confidence intervals (CIs). The end point was revision for any reason or for aseptic loosening of the cup. The revision risks were assessed with use of the Cox multiple regression model. The variables assessed in the Cox model were femoral head size, age group, involved side, operation year, sex, diagnosis, and fixation of the stem. The proportional hazards assumption of the Cox model was not fulfilled, so the follow-up time was divided into 3 time periods: 0 to 2 years, >2 to 4 years, and >4 years. RESULTS: When comparing the 2 groups with regard to revision for any reason, the 5-year Kaplan-Meier survivorship of the Tritanium group (94.7% [95% CI, 94.0% to 95.4%]) was inferior to that of the control group (96.0% [95% CI, 95.7% to 96.3%]). In the Cox regression analysis of the 2 groups for the time period of >4 years, the Tritanium group had an increased risk of revision for any reason compared with the control group (hazard ratio [HR], 3.12 [95% CI, 1.82 to 5.35]; p < 0.001). With regard to revision for aseptic loosening of the cup, the Tritanium group had an increased risk of revision compared with the control group for both 0 to 2 years (HR, 3.80 [95% CI, 1.76 to 8.24]; p < 0.001) and >2 to 4 years (HR, 11.2 [95% CI, 3.28 to 38.0]; p < 0.001). CONCLUSIONS: There was no advantage to using the ultraporous-coated Tritanium cup for primary THA compared with conventional uncemented cups. However, wide CIs for some HR estimates may point to a lack of precision. Therefore, further research on subject is needed. LEVEL OF EVIDENCE: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.


Subject(s)
Arthroplasty, Replacement, Hip , Hip Joint/surgery , Hip Prosthesis , Aged , Aged, 80 and over , Female , Finland , Humans , Male , Middle Aged , Osseointegration , Prosthesis Failure , Registries
4.
Acta Orthop ; 90(4): 312-317, 2019 08.
Article in English | MEDLINE | ID: mdl-30994043

ABSTRACT

Background and purpose - The use of trabecular metal (TM) cups for primary total hip arthroplasty (THA) is increasing. Some recent data suggest that the use of TM in primary THA might be associated with an increased risk of revision. We compared implant survival of Continuum acetabular cups with other commonly used uncemented cups. Patients and methods - Data on 11,390 primary THAs with the Continuum cup and 30,372 THAs with other uncemented cups (reference group) were collected from the Finnish Arthroplasty Register. Kaplan-Meier survival estimates were calculated; the endpoint was revision for any reason, for infection, or for dislocation. Revision risks were assessed with adjusted Cox multiple regression models. A subgroup analysis on the use of neutral or elevated liners in the Continuum group was made. Results - The 7-year survivorship of the Continuum group was 94.6% (95% CI 94.0-95.2) versus 95.6% (CI 95.3-95.8) in the reference group for revision for any reason. The risk for revision was higher in the Continuum group than in the reference group both for revision for any reason (HR 1.3 [CI 1.2-1.5)]) and for revision for dislocation (HR 1.9 [CI 1.5-2.3]). There was no difference in the rates of revision because of infection (HR 0.99 [CI 0.78-1.3]). Use of a neutral liner increased the risk for revision due to dislocation in comparison with the use of an elevated rim liner in the Continuum group (HR 1.7 [CI 1.2-2.5]). Interpretation - THA with Continuum cups is associated with an increased risk of revision compared with other uncemented cups, mainly due to revisions because of dislocation. Our results support the use of an elevated liner when Continuum cups are used for primary THA.


Subject(s)
Arthroplasty, Replacement, Hip/statistics & numerical data , Hip Prosthesis/statistics & numerical data , Prosthesis Failure , Aged , Arthroplasty, Replacement, Hip/adverse effects , Female , Finland/epidemiology , Hip Prosthesis/adverse effects , Humans , Kaplan-Meier Estimate , Male , Registries/statistics & numerical data , Reoperation/statistics & numerical data , Time Factors