Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 8 de 8
Filter
Add more filters










Database
Language
Publication year range
1.
Article in English | MEDLINE | ID: mdl-38082570

ABSTRACT

There is a paucity of data regarding the safety of magnetic resonance imaging (MRI) in patients with abandoned epicardial leads. Few studies have reported temperature rises up to 76 °C during MRI at 1.5 T in gel phantoms implanted with epicardial leads; however, lead trajectories used in these experiments were not clinically relevant. This work reports patient-specific RF heating of both capped and uncapped abandoned epicardial lead configurations during MRI at both 1.5 T and 3 T field strengths. We found that leads routed along realistic, patient-derived trajectories generated substantially lower RF heating than the previously reported worst-case phantom experiments. We also found that MRI at the head imaging landmark leads to substantially lower RF heating compared to MRI at the chest or abdomen landmarks at both 1.5 T and 3 T. Our results suggest that patients with abandoned epicardial leads may safely undergo MRI for head imaging, but caution is warranted during chest and abdominal imaging.Clinical Relevance- Patients with abandoned epicardial leads may safely undergo MRI for head imaging, but caution is warranted during chest and abdominal imaging.


Subject(s)
Heating , Prostheses and Implants , Humans , Phantoms, Imaging , Temperature , Magnetic Resonance Imaging/adverse effects , Magnetic Resonance Imaging/methods
2.
Stereotact Funct Neurosurg ; 101(5): 338-347, 2023.
Article in English | MEDLINE | ID: mdl-37717576

ABSTRACT

INTRODUCTION: Directional deep brain stimulation (DBS) leads have become widely used in the past decade. Understanding the asymmetric stimulation provided by directional leads requires precise knowledge of the exact orientation of the lead in respect to its anatomical target. Recently, the DiODe algorithm was developed to automatically determine the orientation angle of leads from the artifact on postoperative computed tomography (CT) images. However, manual DiODe results are user-dependent. This study analyzed the extent of lead rotation as well as the user agreement of DiODe calculations across the two most common DBS systems, namely, Boston Scientific's Vercise and Abbott's Infinity, and two independent medical institutions. METHODS: Data from 104 patients who underwent an anterior-facing unilateral/bilateral directional DBS implantation at either Northwestern Memorial Hospital (NMH) or Albany Medical Center (AMC) were retrospectively analyzed. Actual orientations of the implanted leads were independently calculated by three individual users using the DiODe algorithm in Lead-DBS and patients' postoperative CT images. The deviation from the intended orientation and user agreement were assessed. RESULTS: All leads significantly deviated from the intended 0° orientation (p < 0.001), regardless of DBS lead design (p < 0.05) or institution (p < 0.05). However, the Boston Scientific leads showed an implantation bias toward the left at both institutions (p = 0.014 at NMH, p = 0.029 at AMC). A difference of 10° between at least two users occurred in 28% (NMH) and 39% (AMC) of all Boston Scientific and 76% (NMH) and 53% (AMC) of all Abbott leads. CONCLUSION: Our results show that there is a significant lead rotation from the intended surgical orientation across both DBS systems and both medical institutions; however, a bias toward a single direction was only seen in the Boston Scientific leads. Additionally, these results raise questions into the user error that occurs when manually refining the orientation angles calculated with DiODe.


Subject(s)
Deep Brain Stimulation , Humans , Retrospective Studies , Deep Brain Stimulation/methods , Electrodes, Implanted , Tomography, X-Ray Computed/methods , Algorithms
3.
Diagnostics (Basel) ; 13(17)2023 Sep 02.
Article in English | MEDLINE | ID: mdl-37685385

ABSTRACT

This study focused on the potential risks of radiofrequency-induced heating of cardiac implantable electronic devices (CIEDs) in children and adults with epicardial and endocardial leads of varying lengths during cardiothoracic MRI scans. Infants and young children are the primary recipients of epicardial CIEDs, though the devices have not been approved as MR conditional by the FDA due to limited data, leading to pediatric hospitals either refusing the MRI service to most pediatric CIED patients or adopting a scan-all strategy based on results from adult studies. The study argues that risk-benefit decisions should be made on an individual basis. We used 120 clinically relevant epicardial and endocardial device configurations in adult and pediatric anthropomorphic phantoms to determine the temperature rise during RF exposure at 1.5 T. The results showed that there was significantly higher RF heating of epicardial leads than endocardial leads in the pediatric phantom, but not in the adult phantom. Additionally, body size and lead length significantly affected RF heating, with RF heating up to 12 °C observed in models based on younger children with short epicardial leads. The study provides evidence-based knowledge on RF-induced heating of CIEDs and highlights the importance of making individual risk-benefit decisions when assessing the potential risks of MRI scans in pediatric CIED patients.

4.
Stereotact Funct Neurosurg ; 101(1): 47-59, 2023.
Article in English | MEDLINE | ID: mdl-36529124

ABSTRACT

INTRODUCTION: Deep brain stimulation (DBS) is a common treatment for a variety of neurological and psychiatric disorders. Recent studies have highlighted the role of neuroimaging in localizing the position of electrode contacts relative to target brain areas in order to optimize DBS programming. Among different imaging methods, postoperative magnetic resonance imaging (MRI) has been widely used for DBS electrode localization; however, the geometrical distortion induced by the lead limits its accuracy. In this work, we investigated to what degree the difference between the actual location of the lead's tip and the location of the tip estimated from the MRI artifact varies depending on the MRI sequence parameters such as acquisition plane and phase encoding direction, as well as the lead's extracranial configuration. Accordingly, an imaging technique to increase the accuracy of lead localization was devised and discussed. METHODS: We designed and constructed an anthropomorphic phantom with an implanted DBS system following 18 clinically relevant configurations. The phantom was scanned at a Siemens 1.5 Tesla Aera scanner using a T1MPRAGE sequence optimized for clinical use and a T1TSE sequence optimized for research purposes. We varied slice acquisition plane and phase encoding direction and calculated the distance between the caudal tip of the DBS lead MRI artifact and the actual tip of the lead, as estimated from MRI reference markers. RESULTS: Imaging parameters and lead configuration substantially altered the difference in the depth of the lead within its MRI artifact on the scale of several millimeters - with a difference as large as 4.99 mm. The actual tip of the DBS lead was found to be consistently more rostral than the tip estimated from the MR image artifact. The smallest difference between the tip of the DBS lead and the tip of the MRI artifact using the clinically relevant sequence (i.e., T1MPRAGE) was found with the sagittal acquisition plane and anterior-posterior phase encoding direction. DISCUSSION/CONCLUSION: The actual tip of an implanted DBS lead is located up to several millimeters rostral to the tip of the lead's artifact on postoperative MR images. This distance depends on the MRI sequence parameters and the DBS system's extracranial trajectory. MRI parameters may be altered to improve this localization.


Subject(s)
Deep Brain Stimulation , Humans , Deep Brain Stimulation/methods , Artifacts , Electrodes, Implanted , Magnetic Resonance Imaging/methods , Brain/pathology
5.
Annu Int Conf IEEE Eng Med Biol Soc ; 2022: 1725-1728, 2022 07.
Article in English | MEDLINE | ID: mdl-36086443

ABSTRACT

Deep brain stimulation (DBS) offers therapeutic benefits to patients suffering from a variety of treatment-resistant neurological and psychiatric disorders. The newest generation of DBS devices now offer directional leads, which utilize segmented electrodes to direct current asymmetrically to the neuronal tissue. Since segmented electrodes offer a larger degree of freedom for contact positioning, it is critical to assess how well the surgically intended and the actual orientation of the lead match to facilitate programming and allow appropriate interpretation of the therapeutic outcome. Postoperative image analysis algorithms, such as DiODe, are commonly used to determine DBS leads' actual orientation. In this work, we used DiODe to compare the deviation between intended and actual orientations of DBS leads across two most commonly implanted directional DBS systems, namely, Boston Scientific Cartesia™ and St. Jude Medical Infinity. This study is the first to investigate the rotation of leads from both DBS systems in a large group of 86 patients. Clinical Relevance- Our results quantify the variability between the surgically intended and actual orientations of Boston Scientific Vercise and St. Jude Medical Infinity DBS systems thus highlighting the need to develop more precise implantation procedures.


Subject(s)
Deep Brain Stimulation , Algorithms , Electrodes , Humans , Rotation
6.
Annu Int Conf IEEE Eng Med Biol Soc ; 2022: 1863-1866, 2022 07.
Article in English | MEDLINE | ID: mdl-36086639

ABSTRACT

Deep brain stimulation (DBS) is an established yet growing treatment for a range of neurological and psychiatric disorders. Over the last decade, numerous studies have underscored the effect of electrode placement on the clinical outcome of DBS. As a result, imaging is now extensively used for DBS electrode localization, even though the accuracy of different modalities in determining the true coordinates of DBS electrodes is less explored. Postoperative magnetic resonance imaging (MRI) is a gold standard method for DBS electrode localization, however, the geometrical distortion induced by the lead's artifact could limit the accuracy. In this work, we investigated to what degree the difference between the true location of the lead's tip and the location of the tip estimated from the MRI artifact varies depending on the MRI sequence parameters, acquisition plane, phase encoding direction, and the implant"s extracranial trajectory. Clinical Relevance- Results will help researchers and clinicians to estimate the true location of DBS leads and contacts from postoperative MRI scans.


Subject(s)
Deep Brain Stimulation , Deep Brain Stimulation/methods , Electrodes, Implanted , Humans , Magnetic Resonance Imaging/methods , Phantoms, Imaging , Postoperative Period
7.
Cardiovasc Eng Technol ; 13(6): 886-898, 2022 12.
Article in English | MEDLINE | ID: mdl-35545752

ABSTRACT

PURPOSE: Peripheral venous pressure (PVP) waveform analysis is a novel, minimally invasive, and inexpensive method of measuring intravascular volume changes. A porcine cohort was studied to determine how venous and arterial pressure waveforms change due to inhaled and infused anesthetics and acute hemorrhage. METHODS: Venous and arterial pressure waveforms were continuously collected, while each pig was under general anesthesia, by inserting Millar catheters into a neighboring peripheral artery and vein. The anesthetic was varied from inhaled to infused, then the pig underwent a controlled hemorrhage. Pearson correlation coefficients between the power of the venous and arterial pressure waveforms at each pig's heart rate frequency were calculated for each variation in the anesthetic, as well as before and after hemorrhage. An analysis of variance (ANOVA) test was computed to determine the significance in changes of the venous pressure waveform means caused by each variation. RESULTS: The Pearson correlation coefficients between venous and arterial waveforms decreased as anesthetic dosage increased. In an opposing fashion, the correlation coefficients increased as hemorrhage occurred. CONCLUSION: Anesthetics and hemorrhage alter venous pressure waveforms in distinctly different ways, making it critical for researchers and clinicians to consider these confounding variables when utilizing pressure waveforms. Further work needs to be done to determine how best to integrate PVP waveforms into clinical decision-making.


Subject(s)
Anesthesia , Arterial Pressure , Swine , Animals , Venous Pressure , Arteries , Hemorrhage/chemically induced , Blood Pressure
8.
J Clin Monit Comput ; 36(1): 147-159, 2022 02.
Article in English | MEDLINE | ID: mdl-33606187

ABSTRACT

Analysis of peripheral venous pressure (PVP) waveforms is a novel method of monitoring intravascular volume. Two pediatric cohorts were studied to test the effect of anesthetic agents on the PVP waveform and cross-talk between peripheral veins and arteries: (1) dehydration setting in a pyloromyotomy using the infused anesthetic propofol and (2) hemorrhage setting during elective surgery for craniosynostosis with the inhaled anesthetic isoflurane. PVP waveforms were collected from 39 patients that received propofol and 9 that received isoflurane. A multiple analysis of variance test determined if anesthetics influence the PVP waveform. A prediction system was built using k-nearest neighbor (k-NN) to distinguish between: (1) PVP waveforms with and without propofol and (2) different minimum alveolar concentration (MAC) groups of isoflurane. 52 porcine, 5 propofol, and 7 isoflurane subjects were used to determine the cross-talk between veins and arteries at the heart and respiratory rate frequency during: (a) during and after bleeding with constant anesthesia, (b) before and after propofol, and (c) at each MAC value. PVP waveforms are influenced by anesthetics, determined by MANOVA: p value < 0.01, η2 = 0.478 for hypovolemic, and η2 = 0.388 for euvolemic conditions. The k-NN prediction models had 82% and 77% accuracy for detecting propofol and MAC, respectively. The cross-talk relationship at each stage was: (a) ρ = 0.95, (b) ρ = 0.96, and (c) could not be evaluated using this cohort. Future research should consider anesthetic agents when analyzing PVP waveforms developing future clinical monitoring technology that uses PVP.


Subject(s)
Anesthetics, Inhalation , Anesthetics , Isoflurane , Propofol , Anesthetics/pharmacology , Animals , Arterial Pressure , Child , Humans , Swine , Venous Pressure
SELECTION OF CITATIONS
SEARCH DETAIL
...