ABSTRACT
OBJECTIVES: The new immunochemistry cobas e 801 module (Roche Diagnostics) was developed to meet increasing demands on routine laboratories to further improve testing efficiency, while maintaining high quality and reliable data. DESIGN AND METHODS: During a non-interventional multicenter evaluation study, the overall performance, functionality and reliability of the new module was investigated under routine-like conditions. It was tested as a dedicated immunochemistry system at four sites and as a consolidator combined with clinical chemistry at three sites. RESULTS: We report on testing efficiency and analytical performance of the new module. Evaluation of sample workloads with site-specific routine request patterns demonstrated increased speed and almost doubled throughput (maximal 300 tests per h), thus revealing that one cobas e 801 module can replace two cobas e 602 modules while saving up to 44% floor space. Result stability was demonstrated by QC analysis per assay throughout the study. Precision testing over 21â¯days yielded excellent results within and between labs, and, method comparison performed versus the cobas e 602 module routine results showed high consistency of results for all assays under study. In a practicability assessment related to performance and handling, 99% of graded features met (44%) or even exceeded (55%) laboratory expectations, with enhanced reagent management and loading during operation being highlighted. CONCLUSION: By nearly doubling immunochemistry testing efficiency on the same footprint as a cobas e 602 module, the new module has a great potential to further consolidate and enhance laboratory testing while maintaining high quality analytical performance with Roche platforms.