Opt Express ; 26(23): 30523-30531, 2018 Nov 12.
Article in English | MEDLINE | ID: mdl-30469950

ABSTRACT

We demonstrate an optically pumped magnetometer (OPM) operated in a free-induction-decay (FID) configuration that is capable of tracking oscillating magnetic signals in the presence of a 50 µT static field. Excellent waveform reconstruction is demonstrated for low frequency modulations with respect to the Nyquist limited bandwidth. A 100 pT oscillation was successfully reconstructed using signal averaging, and an optimum sensitivity of 3.9 pT/Hz was measured from the spectrum of the residuals relative to the sinusoidal fit. The impact of the pump-probe repetition rate and spin depolarization on the frequency response of the sensor is investigated in detail using miniaturized vapor cell technology, with the (-3 dB) bandwidths residing beyond the Nyquist limit in each case. We also discuss technical limitations associated with the magnetometer when exposed to oscillating fields of sufficiently high amplitude or frequency. This is discussed in the context of potential distortions arising in the reproduced signals, induced by frequency modulation (FM) and aliasing artefacts.