Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
1.
J Biomed Mater Res ; 48(3): 251-7, 1999.
Article in English | MEDLINE | ID: mdl-10398028

ABSTRACT

Uniaxial fatigue testing was performed on different diameters of fine wires made from MP35N. The fatigue limits of the wires differed from each other based on the diameter of the wire. Multiaxial (shear) fatigue testing was also performed on a benchmark coil used to evaluate the fatigue life of all modern pacemaker leads (the CENELEC standard coil). A computer algorithm was used to quantify the maximum shear stress and strain on the coil. The bend radius, coil diameter, wire diameter, and pitch of the coil all affect the shear stress and strain and therefore the fatigue properties of conductor coils. Based on the analysis presented, it was determined that the portion of the CENELEC standard dealing with fatigue, when used in its present format, is not a valid fatigue test for pacemaker leads.


Subject(s)
Biocompatible Materials , Lead , Pacemaker, Artificial , Humans , Mechanics