Int J Otolaryngol ; 2017: 2707690, 2017.
Article in English | MEDLINE | ID: mdl-28553354

ABSTRACT

Objective. To assess the face, content, and construct validity of the Voxel-Man TempoSurg Virtual Reality simulator. Participants and Methods. 74 ear, nose, and throat (ENT) surgeons participated. They were assigned to one of two groups according to their level of expertise: the expert group (n = 16) and the novice group (n = 58). The participants performed four temporal bone dissection tasks on the simulator. Performances were assessed by a global score and then compared to assess the construct validity of the simulator. Finally, the expert group assessed the face and content validity by means of a five-point Likert-type scale. Results. experienced surgeons performed better (p < .01) and faster (p < .001) than the novices. However, the groups did not differ in terms of bone volume removed (p = .11) or number of injuries (p = .37). 93.7% of experienced surgeons stated they would recommend this simulator for anatomical learning. Most (87.5%) also thought that it could be integrated into surgical training. Conclusion. The Voxel-Man TempoSurg Virtual Reality simulator constitutes an interesting complementary tool to traditional teaching methods for training in otologic surgery.