Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters











Database
Language
Publication year range
1.
Perfusion ; 21(2): 83-6, 2006 Mar.
Article in English | MEDLINE | ID: mdl-16615684

ABSTRACT

INTRODUCTION: Microemboli are the main implicated cause of neuropsychological (NP) impairment after cardiac surgery. This prospective clinical trial compared the effect of an auto-venting arterial line filter on intraoperative cerebral microemboli and NP outcome compared to an arterial line filter with a vent line, in patients undergoing elective coronary artery bypass graft (CABG) surgery. METHODS: One hundred and ten patients received either an Avecor Affinity (n =73) or Pall AV-6 (n =37) control filter. Cerebral microemboli during cardiopulmonary bypass were recorded by transcranial Doppler monitoring of the right middle cerebral artery. Evidence of cerebral impairment was obtained by comparing patients' performance in a NP test battery (nine tests) administered 6-8 weeks postoperatively with their preoperative scores. RESULTS: During cardiopulmonary bypass, the median number and range of microemboli were 67 (5-846) and 55 (2-773) for the Avecor and AV-6 groups, respectively (p = 0.47). There was no difference in NP outcome. CONCLUSION: There is no difference in the filtering ability of vent-line and auto-vent filters as assessed by cerebral microemboli. This, together with the similar NP outcome, suggests that both types of filter are equally safe for clinical use.


Subject(s)
Extracorporeal Circulation/instrumentation , Intracranial Embolism/prevention & control , Aged , Extracorporeal Circulation/adverse effects , Female , Filtration/instrumentation , Humans , Intracranial Embolism/diagnosis , Intracranial Embolism/diagnostic imaging , Male , Middle Aged , Middle Cerebral Artery/diagnostic imaging , Neuropsychological Tests , Prospective Studies , Treatment Outcome , Ultrasonography, Doppler, Transcranial/methods
2.
Eur J Cardiothorac Surg ; 25(2): 267-74, 2004 Feb.
Article in English | MEDLINE | ID: mdl-14747125

ABSTRACT

OBJECTIVES: A randomised clinical trial sought evidence as to whether leucocyte-depleting (LD) arterial line filters added a further degree of neuroprotection in patients undergoing elective coronary artery bypass graft (CABG) surgery. METHODS: One hundred and ninety-two patients were randomised to the use of a Pall Leukoguard-6 LD filter or either an Avecor Affinity or Pall Autovent-6 control filter. Cerebral microemboli during surgery were recorded by transcranial Doppler (TCD) monitor over the right middle cerebral artery. Evidence of cerebral impairment was obtained by comparing patients' performance in a neuropsychological (NP) test battery (nine tests) administered 6-8 weeks post-operatively with their pre-operative scores. RESULTS: The groups proved well balanced in pre-operative variables. During cardiopulmonary bypass (CPB) the median number and range of microemboli was 15 (3-180) in the LD group compared to 67 (5-846) and 55 (2-773) for the Avecor and AV6 groups, respectively (P<0.0001). One hundred and sixty-two patients completed all the NP tests. The LD group showed better post-operative performance in all but one of the nine tests although the difference in a total change score just failed to reach significance (P=0.07 one-tailed t-test). CONCLUSIONS: LD filtration during CABG reduced the number of cerebral microemboli recorded by TCD and showed a strong trend towards improving NP performance post-operatively. These findings suggest that the use of such filters in CABG surgery may offer increased neuroprotection.


Subject(s)
Cognition Disorders/prevention & control , Coronary Artery Bypass/adverse effects , Intracranial Embolism and Thrombosis/prevention & control , Intraoperative Care/methods , Leukapheresis/methods , Aged , Blood Flow Velocity , Cognition Disorders/etiology , Female , Filtration , Humans , Intracranial Embolism and Thrombosis/etiology , Leukocyte Count , Male , Middle Aged , Middle Cerebral Artery/diagnostic imaging , Middle Cerebral Artery/physiopathology , Neuropsychological Tests , Prospective Studies , Ultrasonography
SELECTION OF CITATIONS
SEARCH DETAIL