Radiat Prot Dosimetry ; 146(1-3): 58-61, 2011 Jul.
Article in English | MEDLINE | ID: mdl-21493614

ABSTRACT

The main research described in this paper includes three sections. First, research on the response of the stainless steel ball-shaped ionisation chamber by experimental methods. Secondly, calculation of the response of the chamber with the general Monte Carlo EGS4 code in order to compare with the equivalent electron source theory by calculation methods. Finally, calculation of the response of the ionisation chamber with the equivalent electron source theory. The results show that the calculated results of the equivalent electron source theory coincide very well with those of the experiments when the atomic number of the chamber wall is close to one of the gases (such as Ar and Kr), and the calculated results coincide with those of the experiments to a certain extent when the atomic number of the chamber wall is not close to one of the gases (such as He and Xe).


Subject(s)
Algorithms , Computer Simulation , Electrons , Monte Carlo Method , Radiation, Ionizing , Radiometry/instrumentation , Calibration , Scattering, Radiation