J Extra Corpor Technol ; 22(2): 61-6, 1990.
Article in English | MEDLINE | ID: mdl-10171101

ABSTRACT

Three devices used to measure hemoglobin oxygen saturation in the extracorporeal circuit were studied and compared to a control. The Baxter Bentley OxySat, Oximetrix Accusat, and Radiometer ABL4 blood gas monitor were compared to a control, the IL 282 Co-Oximeter. Fifty-one sample points were obtained during all phases of cardiopulmonary bypass with results as follows: table: see text. The Accusat was found to be a statistically more accurate means of monitoring hemoglobin oxygen saturations during cardiopulmonary bypass than the ABL4 and the OxySat. All devices had significant correlation with the control and with each other.


Subject(s)
Cardiopulmonary Bypass , Extracorporeal Circulation/instrumentation , Hemoglobins/analysis , Monitoring, Intraoperative/instrumentation , Oxygen/blood , Adult , Female , Humans , Male