1.
Analyst ; 138(13): 3621-5, 2013 Jul 07.
Article in English | MEDLINE | ID: mdl-23702824

ABSTRACT

We report an improved interface for two-dimensional capillary electrophoresis. This interface is based on capillary tubing and a Plexiglas chip, both of which were milled using a micro-dicing saw. The interface was evaluated and compared to a traditional interface design for both pseudo one-dimensional and two-dimensional capillary electrophoresis. We observe less than 70% transfer efficiency for the traditional design and greater than 90% transfer efficiency with this new interface.


Subject(s)
Electrophoresis, Capillary/instrumentation , Equipment Design , Microtechnology
2.
Analyst ; 137(13): 3099-101, 2012 Jul 07.
Article in English | MEDLINE | ID: mdl-22606689

ABSTRACT

We present a design for a sheath-flow cuvette that uses a relatively inexpensive quartz cuvette. The cuvette has a high optical quality square flow chamber that is fused to quartz tubes at each end. PEEK/TEFZEL fittings hold and seal the quartz flow chamber without putting strain on the cuvette. The performance of the cuvette is evaluated as a laser-induced fluorescence detector for capillary electrophoresis. The cuvette produces mass detection limits of 50 yoctomoles (30 copies) for 5-carboxyl tetramethylrhodamine (5 TAMRA SE) with a separation efficiency of 400,000 theoretical plates.


Subject(s)
Electrophoresis, Capillary/methods , Spectrometry, Fluorescence/methods , Flow Cytometry , Limit of Detection