Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Language
Publication year range
1.
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-589124

ABSTRACT

18 years of age were enrolled. The blood pressure measurements at the wrist were compared with the results obtained by auscultatory measurement on left upper arm. Both measurements were taken 3 times in each person. A total 273 measurements were obtained in 91 participants. The evaluation of the accuracy was assessed by 2 statistical methods. (1)The differences between the 2 devices for a given measurement. (2) The differences between the 2 devices for a given person. Results Method 1 showed the sample mean errors and standard deviation of errors were 0.5?7.1 mm Hg for systolic blood pressure(SBP) and -1.6?6.1 mm Hg for diastolic blood pressure(DBP). Method 2 showed they were 0.5?6.1 mm Hg for SBP and -1.6?5.3 mm Hg for DBP. Blood pressure level and the wrist circumferences had no influence on the measurement errors. Conclusion The HEM-6000 wrist blood pressure device apparently met the AAMI criteria and is valid for patient self-measurement.

SELECTION OF CITATIONS
SEARCH DETAIL