2.
Scand J Clin Lab Invest ; 47(2): 191-3, 1987 Apr.
Article in English | MEDLINE | ID: mdl-3576122

ABSTRACT

In 92 consecutive patients admitted with suspected acute deep venous thrombosis, (DVT), the screening value of a new thermographic equipment, DeVeTherm, was evaluated with phlebography as a reference. The predictive value was 0.29 (0.19-0.41) of a positive thermography, and 0.68 (0.43-0.87) of a negative thermography. These results do not justify application of DeVeTherm as a screening method for DVT.


Subject(s)
Body Temperature , Leg/blood supply , Thermography , Thrombophlebitis/diagnosis , Acute Disease , Adult , Aged , Evaluation Studies as Topic , Female , Humans , Male , Middle Aged , Phlebography , Thrombophlebitis/physiopathology