Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 3 de 3
Filter
Add more filters










Database
Language
Publication year range
1.
PLoS One ; 18(2): e0278561, 2023.
Article in English | MEDLINE | ID: mdl-36758026

ABSTRACT

INTRODUCTION: The newly introduced Cellular Oxygen METabolism (COMET®) monitor enables the measurement of mitochondrial oxygen tension (mitoPO2) using the protoporphyrin IX triplet state lifetime technique (PpIX-TSLT). This study aims to investigate the feasibility and applicability of the COMET® measurements in the operating theatre and study the behavior of the new parameter mitoPO2 during stable operating conditions. METHODS: In this observational study mitochondrial oxygenation was measured in 20 patients during neurosurgical procedures using the COMET® device. Tissue oxygenation and local blood flow were measured by the Oxygen to See (O2C). Primary outcomes included mitoPO2, skin temperature, mean arterial blood pressure, local blood flow and tissue oxygenation. RESULTS: All patients remained hemodynamically stable during surgery. Mean baseline mitoPO2 was 60 ± 19 mmHg (mean ± SD) and mean mitoPO2 remained between 40-60 mmHg during surgery, but tended to decrease over time in line with increasing skin temperature. CONCLUSION: This study presents the feasibility of mitochondrial oxygenation measurements as measured by the COMET® monitor in the operating theatre and shows the parameter mitoPO2 to behave in a stable and predictable way in the absence of notable hemodynamic alterations. The results provide a solid base for further research into the added value of mitochondrial oxygenation measurements in the perioperative trajectory.


Subject(s)
Mitochondria , Oxygen , Humans , Mitochondria/metabolism , Blood Gas Analysis , Oxygen/metabolism , Oxygen Consumption , Respiratory Physiological Phenomena
2.
J Clin Monit Comput ; 31(6): 1143-1150, 2017 Dec.
Article in English | MEDLINE | ID: mdl-28000040

ABSTRACT

After introduction of the protoporphyrin IX-triplet state lifetime technique as a new method to measure mitochondrial oxygen tension in vivo, the development of a clinical monitor was started. This monitor is the "COMET", an acronym for Cellular Oxygen METabolism. The COMET is a non-invasive electrically powered optical device that allows measurements on the skin. The COMET is easy to transport, due to its lightweight and compact size. After 5-aminolevulinic acid application on the human skin, a biocompatible sensor enables detection of PpIX in the mitochondria. PpIX acts as a mitochondrially located oxygen-sensitive dye. Three measurement types are available in the touchscreen-integrated user interface, 'Single', 'Interval' and 'Dynamic measurement'. COMET is currently used in several clinical studies in our institution. In this first description of the COMET device we show an incidental finding during neurosurgery. To treat persisting intraoperative hypertension a patient was administered clonidine, but due to rapid administration an initial phase of peripheral vasoconstriction occurred. Microvascular flow and velocity parameters measured with laser-doppler (O2C, LEA Medizintechnik) decreased by 44 and 16% respectively, but not the venous-capillary oxygen saturation. However, mitochondrial oxygen tension in the skin detected by COMET decreased from a steady state of 48 to 16 mmHg along with the decrease in flow and velocity. We conclude that COMET is ready for clinical application and we see the future for this bedside monitor on the intensive care, operating theater, and testing of mitochondrial effect of pharmaceuticals.


Subject(s)
Biocompatible Materials/chemistry , Mitochondria/metabolism , Monitoring, Ambulatory/methods , Oxygen/metabolism , Protoporphyrins/chemistry , Skin/pathology , Aminolevulinic Acid/chemistry , Equipment Design , Humans , Lasers , Microcirculation , Oxygen Consumption , Signal Processing, Computer-Assisted , Skin/metabolism
SELECTION OF CITATIONS
SEARCH DETAIL
...