1.
Med Phys ; 42(9): 5270-7, 2015 Sep.
Article in English | MEDLINE | ID: mdl-26328976

ABSTRACT

PURPOSE: For a given bowtie filter design, both the selection of material and the physical design control the energy fluence, and consequently the dose distribution, in the object. Using three previously described bowtie filter designs, the goal of this work is to demonstrate the effect that different materials have on the bowtie filter performance measures. METHODS: Three bowtie filter designs that compensate for one or more aspects of the beam-modifying effects due to the differences in path length in a projection have been designed. The nature of the designs allows for their realization using a variety of materials. The designs were based on a phantom, 14 cm in diameter, composed of 40% fibroglandular and 60% adipose tissue. Bowtie design #1 is based on single material spectral matching and produces nearly uniform spectral shape for radiation incident upon the detector. Bowtie design #2 uses the idea of basis-material decomposition to produce the same spectral shape and intensity at the detector, using two different materials. With bowtie design #3, it is possible to eliminate the beam hardening effect in the reconstructed image by adjusting the bowtie filter thickness so that the effective attenuation coefficient for every ray is the same. Seven different materials were chosen to represent a range of chemical compositions and densities. After calculation of construction parameters for each bowtie filter design, a bowtie filter was created using each of these materials (assuming reasonable construction parameters were obtained), resulting in a total of 26 bowtie filters modeled analytically and in the penelope Monte Carlo simulation environment. Using the analytical model of each bowtie filter, design profiles were obtained and energy fluence as a function of fan-angle was calculated. Projection images with and without each bowtie filter design were also generated using penelope and reconstructed using FBP. Parameters such as dose distribution, noise uniformity, and scatter were investigated. RESULTS: Analytical calculations with and without each bowtie filter show that some materials for a given design produce bowtie filters that are too large for implementation in breast CT scanners or too small to accurately manufacture. Results also demonstrate the ability to manipulate the energy fluence distribution (dynamic range) by using different materials, or different combinations of materials, for a given bowtie filter design. This feature is especially advantageous when using photon counting detector technology. Monte Carlo simulation results from penelope show that all studied material choices for bowtie design #2 achieve nearly uniform dose distribution, noise uniformity index less than 5%, and nearly uniform scatter-to-primary ratio. These same features can also be obtained using certain materials with bowtie designs #1 and #3. CONCLUSIONS: With the three bowtie filter designs used in this work, the selection of material is an important design consideration. An appropriate material choice can improve image quality, dose uniformity, and dynamic range.


Subject(s)
Image Processing, Computer-Assisted/methods , Mammography/methods , Tomography, X-Ray Computed/methods , Radiation Dosage , Scattering, Radiation , Signal-To-Noise Ratio
2.
Med Phys ; 42(3): 1453-62, 2015 Mar.
Article in English | MEDLINE | ID: mdl-25735298

ABSTRACT

PURPOSE: To design bowtie filters with improved properties for dedicated breast CT to improve image quality and reduce dose to the patient. METHODS: The authors present three different bowtie filters designed for a cylindrical 14-cm diameter phantom with a uniform composition of 40/60 breast tissue, which vary in their design objectives and performance improvements. Bowtie design #1 is based on single material spectral matching and produces nearly uniform spectral shape for radiation incident upon the detector. Bowtie design #2 uses the idea of basis material decomposition to produce the same spectral shape and intensity at the detector, using two different materials. Bowtie design #3 eliminates the beam hardening effect in the reconstructed image by adjusting the bowtie filter thickness so that the effective attenuation coefficient for every ray is the same. All three designs are obtained using analytical computational methods and linear attenuation coefficients. Thus, the designs do not take into account the effects of scatter. The authors considered this to be a reasonable approach to the filter design problem since the use of Monte Carlo methods would have been computationally intensive. The filter profiles for a cone-angle of 0° were used for the entire length of each filter because the differences between those profiles and the correct cone-beam profiles for the cone angles in our system are very small, and the constant profiles allowed construction of the filters with the facilities available to us. For evaluation of the filters, we used Monte Carlo simulation techniques and the full cone-beam geometry. Images were generated with and without each bowtie filter to analyze the effect on dose distribution, noise uniformity, and contrast-to-noise ratio (CNR) homogeneity. Line profiles through the reconstructed images generated from the simulated projection images were also used as validation for the filter designs. RESULTS: Examples of the three designs are presented. Initial verification of performance of the designs was done using analytical computations of HVL, intensity, and effective attenuation coefficient behind the phantom as a function of fan-angle with a cone-angle of 0°. The performance of the designs depends only weakly on incident spectrum and tissue composition. For all designs, the dynamic range requirement on the detector was reduced compared to the no-bowtie-filter case. Further verification of the filter designs was achieved through analysis of reconstructed images from simulations. Simulation data also showed that the use of our bowtie filters can reduce peripheral dose to the breast by 61% and provide uniform noise and CNR distributions. The bowtie filter design concepts validated in this work were then used to create a computational realization of a 3D anthropomorphic bowtie filter capable of achieving a constant effective attenuation coefficient behind the entire field-of-view of an anthropomorphic breast phantom. CONCLUSIONS: Three different bowtie filter designs that vary in performance improvements were described and evaluated using computational and simulation techniques. Results indicate that the designs are robust against variations in breast diameter, breast composition, and tube voltage, and that the use of these filters can reduce patient dose and improve image quality compared to the no-bowtie-filter case.


Subject(s)
Breast , Mammography , Models, Theoretical , Radiographic Image Enhancement/methods , Humans , Monte Carlo Method , Phantoms, Imaging , Signal-To-Noise Ratio
3.
J Med Imaging (Bellingham) ; 1(3): 033505, 2014 Oct.
Article in English | MEDLINE | ID: mdl-26158057

ABSTRACT

Scatter contamination of projection images in cone-beam computed tomography (CT) degrades the image quality. The use of bowtie filters in dedicated breast CT can decrease this scatter contribution. Three bowtie filter designs that compensate for one or more aspects of the beam-modifying effects due to differences in path length in a projection were studied. These designs have been investigated in terms of their ability to reduce the scatter contamination in projection images acquired in a dedicated breast CT geometry. The scatter magnitude was measured as the scatter-to-primary ratio (SPR) using experimental and Monte Carlo techniques for various breast phantom diameters and tube voltages. The results show that a 55% reduction in the center SPR value could be obtained with the bowtie filter designs. On average, the bowtie filters reduced the center SPR by approximately 18% over all breast diameters. The distribution of the scatter was calculated at a range of different locations to produce scatter distribution maps for all three bowtie filter designs. With the inclusion of the bowtie filters, the scatter distribution was more uniform for all breast diameters. The results of this study will be useful in designing scatter correction methods and understanding the benefits of bowtie filters in dedicated breast CT.

4.
Med Phys ; 40(5): 051914, 2013 May.
Article in English | MEDLINE | ID: mdl-23635284

ABSTRACT

PURPOSE: Digital breast tomosynthesis (DBT) is a promising breast cancer screening tool that has already begun making inroads into clinical practice. However, there is ongoing debate over how to quantitatively evaluate and optimize these systems, because different definitions of image quality can lead to different optimal design strategies. Powerful and accurate tools are desired to extend our understanding of DBT system optimization and validate published design principles. METHODS: The authors developed a virtual trial framework for task-specific DBT assessment that uses digital phantoms, open-source x-ray transport codes, and a projection-space, spatial-domain observer model for quantitative system evaluation. The authors considered evaluation of reconstruction algorithms as a separate problem and focused on the information content in the raw, unfiltered projection images. Specifically, the authors investigated the effects of scan angle and number of angular projections on detectability of a small (3 mm diameter) signal embedded in randomly-varying anatomical backgrounds. Detectability was measured by the area under the receiver-operating characteristic curve (AUC). Experiments were repeated for three test cases where the detectability-limiting factor was anatomical variability, quantum noise, or electronic noise. The authors also juxtaposed the virtual trial framework with other published studies to illustrate its advantages and disadvantages. RESULTS: The large number of variables in a virtual DBT study make it difficult to directly compare different authors' results, so each result must be interpreted within the context of the specific virtual trial framework. The following results apply to 25% density phantoms with 5.15 cm compressed thickness and 500 µm(3) voxels (larger 500 µm(2) detector pixels were used to avoid voxel-edge artifacts): 1. For raw, unfiltered projection images in the anatomical-variability-limited regime, AUC appeared to remain constant or increase slightly with scan angle. 2. In the same regime, when the authors fixed the scan angle, AUC increased asymptotically with the number of projections. The threshold number of projections for asymptotic AUC performance depended on the scan angle. In the quantum- and electronic-noise dominant regimes, AUC behaviors as a function of scan angle and number of projections sometimes differed from the anatomy-limited regime. For example, with a fixed scan angle, AUC generally decreased with the number of projections in the electronic-noise dominant regime. These results are intended to demonstrate the capabilities of the virtual trial framework, not to be used as optimization rules for DBT. CONCLUSIONS: The authors have demonstrated a novel simulation framework and tools for evaluating DBT systems in an objective, task-specific manner. This framework facilitates further investigation of image quality tradeoffs in DBT.


Subject(s)
Breast Neoplasms/diagnostic imaging , Mammography/methods , Radiographic Image Enhancement/methods , Humans , Phantoms, Imaging , Quality Control , Radiation Dosage
5.
Phys Med Biol ; 56(12): 3513-33, 2011 Jun 21.
Article in English | MEDLINE | ID: mdl-21606556

ABSTRACT

The purpose of this study is to characterize the x-ray properties of a dual-modality, anthropomorphic breast phantom whose MRI properties have been previously evaluated. The goal of this phantom is to provide a platform for optimization and standardization of two- and three-dimensional x-ray and MRI breast imaging modalities for the purpose of lesion detection and discrimination. The phantom is constructed using a mixture of lard and egg whites, resulting in a variable, tissue-mimicking structure with separate adipose- and glandular-mimicking components. The phantom can be produced with either a compressed or uncompressed shape. Mass attenuation coefficients of the phantom materials were estimated using elemental compositions from the USDA National Nutrient Database for Standard Reference and the atomic interaction models from the Monte Carlo code PENELOPE and compared with human values from the literature. The image structure was examined quantitatively by calculating and comparing spatial covariance matrices of the phantom and patient mammography images. Finally, a computerized version of the phantom was created by segmenting a computed tomography scan and used to simulate x-ray scatter of the phantom in a mammography geometry. Mass attenuation coefficients of the phantom materials were within 20% and 15% of the values for adipose and glandular tissues, respectively, which is within the estimation error of these values. Matching was improved at higher energies (>20 keV). Tissue structures in the phantom have a size similar to those in the patient data, but are slightly larger on average. Correlations in the patient data appear to be longer than those in the phantom data in the anterior-posterior direction; however, they are within the error bars of the measurement. Simulated scatter-to-primary ratio values of the phantom images were as high as 85% in some areas and were strongly affected by the heterogeneous nature of the phantom. Key physical x-ray properties of the phantom have been quantitatively evaluated and shown to be comparable to those of breast tissue. Since the MRI properties of the phantom have been previously evaluated, we believe it is a useful tool for quantitative evaluation of two- and three-dimensional x-ray and MRI breast imaging modalities for the purpose of lesion detection and characterization.


Subject(s)
Breast , Magnetic Resonance Imaging/instrumentation , Mammography/instrumentation , Phantoms, Imaging , Breast/pathology , Breast Neoplasms/diagnostic imaging , Breast Neoplasms/pathology , Humans , Reproducibility of Results , Scattering, Radiation , X-Rays
6.
IEEE Trans Med Imaging ; 28(5): 696-702, 2009 May.
Article in English | MEDLINE | ID: mdl-19272986

ABSTRACT

We quantify the variation in resolution due to anisotropy caused by oblique X-ray incidence in indirect flat-panel detectors for computed tomography breast imaging systems. We consider a geometry and detector type utilized in breast computed tomography (CT) systems currently being developed. Our methods rely on mantis, a combined X-ray, electron, and optical Monte Carlo transport open source code. The physics models are the most accurate available in general-purpose Monte Carlo packages in the diagnostic energy range. We consider maximum-obliquity angles of 10 ( degrees ) and 13 ( degrees ) at the centers of the 30 and 40 cm detector edges, respectively, and 16 ( degrees ) at the corner of the detector. Our results indicate that blur is asymmetric and that the resolution properties vary significantly with the angle (or location) of incidence. Our results suggest that the asymmetry can be as high as a factor of 2.6 between orthogonal directions. Anisotropy maps predicted by mantis provide an understanding of the effect that such variations have on the imaging system and allow more accurate modeling and optimization of breast CT systems. These maps of anisotropy across the detector could lead to improved reconstruction and help motivate physics-based strategies for computer detection of breast lesions.


Subject(s)
Image Interpretation, Computer-Assisted/methods , Imaging, Three-Dimensional/methods , Mammography/methods , Tomography, X-Ray Computed/methods , Anisotropy , Cesium , Computer Simulation , Female , Humans , Iodides , Monte Carlo Method , Normal Distribution
7.
Med Phys ; 34(11): 4076-91, 2007 Nov.
Article in English | MEDLINE | ID: mdl-18074617

ABSTRACT

We describe the anisotropy in imaging performance caused by oblique x-ray incidence in indirect detectors for breast tomosynthesis based on columnar scintillator screens. We use MANTIS, a freely available combined x-ray, electron, and optical Monte Carlo transport package which models the indirect detection processes in columnar screens, interaction by interaction. The code has been previously validated against published optical distributions. In this article, initial validation results are provided concerning the blur for particular designs of phosphor screens for which some details with respect to the columnar geometry are available from scanning electron microscopy. The polyenergetic x-ray spectrum utilized comes from a database of experimental data for three different anode/filter/kVp combinations: Mo/Mo at 28 kVp, Rh/Rh at 28 kVp, and W/Al at 42 kVp. The x-ray spectra were then filtered with breast tissue (3, 4, and 6 cm thickness), compression paddle, and support base, according to the oblique paths determined by the incidence angle. The composition of the breast tissue was 50%/50% adipose/glandular tissue mass ratio. Results are reported on the pulse-height statistics of the light output and on spatial blur, expressed as the response of the detector to a pencil beam with a certain incidence angle. Results suggest that the response is nonsymmetrical and that the resolution properties of a tomosynthesis system vary significantly with the angle of x-ray incidence. In contrast, it is found that the noise due to the variability in the number of light photons detected per primary x-ray interaction changes only a few percent. The anisotropy in the response is not less in screens with absorptive backings while the noise introduced by variations in the depth-dependent light output and optical transport is larger. The results suggest that anisotropic imaging performance across the detector area can be incorporated into reconstruction algorithms for improving the image quality of breast tomosynthesis. This study also demonstrates that the assessment of image quality of breast tomosynthesis systems requires a more complete description of the detector response beyond local, center measurements of resolution and noise that assume some degree of symmetry in the detector performance.


Subject(s)
Breast/pathology , Imaging, Three-Dimensional/instrumentation , Mammography/instrumentation , Radiographic Image Enhancement/instrumentation , Radiographic Image Interpretation, Computer-Assisted/instrumentation , X-Ray Intensifying Screens , Adipose Tissue/metabolism , Anisotropy , Cesium/pharmacology , Equipment Design , Female , Humans , Iodides/pharmacology , Monte Carlo Method , Phosphorus/chemistry , Sensitivity and Specificity , X-Rays
8.
Med Phys ; 31(11): 3122-31, 2004 Nov.
Article in English | MEDLINE | ID: mdl-15587665

ABSTRACT

Noise transfer in granular x-ray imaging phosphor screens is not proportional to the square of the magnitude of the signal transfer when the transfer properties are considered for the entire screen thickness, unless appropriately weighted at each depth of interaction. This property, known as the Lubberts effect, has not yet been studied in columnar structured screens because of a lack of a generalized description of the depth-dependent light transport. In this paper, we investigate the signal and noise transfer characteristics of columnar phosphors used in digital mammography detectors using DETECT-II, an optical Monte Carlo light transport simulation code. We first validate our choice of optical parameters for the description of granular and columnar screens using published normalized modulation transfer (MTF) experimental data. Our calculations of MTF match empirically measured MTFs for a granular film/screen analog system, and for an indirect x-ray digital imaging system with CsI:Tl screen representative of digital mammography systems. Using the depth-dependent spread functions and collection efficiencies, we calculate the signal and noise transfer functions and the Lubberts fraction, which is the ratio of the signal transfer function to the noise transfer function, for different screen thicknesses of granular and columnar phosphors. We find that the Lubberts fraction of a 85 microm granular screen model corresponding to a Gd2O2S:Tb screen is similar to the fraction for a 100 microm columnar CsI:Tl screen.


Subject(s)
Equipment Failure Analysis/methods , Models, Theoretical , Phosphorus/radiation effects , Radiographic Image Enhancement/instrumentation , Radiometry/instrumentation , Radiometry/methods , X-Ray Intensifying Screens , Computer Simulation , Radiation Dosage , Radiographic Image Enhancement/methods , Reproducibility of Results , Sensitivity and Specificity
9.
Med Phys ; 31(4): 715-23, 2004 Apr.
Article in English | MEDLINE | ID: mdl-15124988

ABSTRACT

Subpixel structures found in medical monochrome active-matrix liquid crystal displays (AMLCDs) affect noise estimates measured with conventional methods. In this work, we discuss methods that identify sources of noise and permit the comparison of luminance noise estimates across technologies independent of pixel design and device technology. We used a three-million pixel AMLCD with a pixel structure consisting of three color stripes, each in a two-domain, in-plane switching mode. Images of uniform fields displayed on the AMLCD were acquired using a low-noise, high-resolution CCD camera. The camera noise and flat-field response were characterized using a uniform light source constructed for this purpose. We show results in terms of spatial luminance noise and noise power spectrum for high-resolution images and for the same images processed with a pixel-aligned aperture. We find that the pixel-aligned aperture eliminates almost all the noise found in the high-resolution images, suggesting that most of the luminance noise in AMLCDs comes from the subpixel structure and less-than-100% aperture ratio, rather than from interpixel variations.


Subject(s)
Computer Terminals , Data Display , Diagnostic Imaging/instrumentation , Equipment Failure Analysis , Image Interpretation, Computer-Assisted/methods , Signal Processing, Computer-Assisted/instrumentation , Stochastic Processes , Reproducibility of Results , Sensitivity and Specificity