Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
J Trauma ; 46(4): 607-11; discussion 611-2, 1999 Apr.
Article in English | MEDLINE | ID: mdl-10217222

ABSTRACT

BACKGROUND: The pulmonary artery (PA) catheter has been used to determine hemodynamic indices; however, it has recently been criticized. This study was undertaken to evaluate an esophageal Doppler monitor (EDM) as a possible replacement for PA catheter in critically ill, mechanically ventilated patients. METHODS: EDM and PA catheters were placed in patients in the surgical intensive care units (n = 14, 118 matched sets of data). PA catheter and EDM measurements, including corrected flow time (FTc,) a measure of preload, were obtained. Pearson correlation (r) was analyzed to compare PA catheter and EDM measurements, and a nonlinear regression model was used to describe Starling Relationships. RESULTS: Cardiac output correlated between EDM and PA catheter (r = 0.6; p < 0.001). FTc correlated more strongly with cardiac output than did pulmonary capillary wedge pressure. (FTc: r2 = 0.27; p < 0.001; cardiac output: r2 = 0.04; p = 0.06). CONCLUSION: Corrected flow time is a better indicator of preload than pulmonary capillary wedge pressures. EDM seems to be at least as useful as PA catheter in managing the hemodynamic status of critically ill surgical patients.


Subject(s)
Catheterization, Swan-Ganz , Critical Care , Critical Illness/therapy , Esophagus/diagnostic imaging , Hemodynamics , Adult , Aged , Female , Humans , Male , Middle Aged , Monitoring, Physiologic/methods , Postoperative Complications , Ultrasonography
SELECTION OF CITATIONS
SEARCH DETAIL
...