Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
PLoS One ; 18(9): e0291507, 2023.
Article in English | MEDLINE | ID: mdl-37699042

ABSTRACT

This study aimed to investigate the feasibility of applying 3D gel dosimeters for proton therapy. Two different formulations (5-5-3-5, 5-3-3-10) for the N-Isopropyl Acrylamide (NIPAM) polymer gel were used to find the best composition for the application of NIPAM polymer gels for proton therapy. The reaction of the gel under different physical conditions, including dependence on energy and dependence on the dose rate of the NIPAM gel under proton irradiation, was also explored. A NIPAM gel dosimeter was used to record the 3D dose distribution, and a self-developed parallel beam optical computed tomography scanner was used to obtain non-irradiated and post-irradiated gel phantom images. The NIPAM gel was filled into a cylindrical acrylic phantom. The results showed that the optical density of the irradiated NIPAM dosimeter was linear in the dose range of 0 to 6 Gy, and the linearity of the two NIPAM gel formulations at the depth of the dose point (2 cm) was 0.98 to 0.89. The dose depth curves showed different patterns with different gel sensitivities. This study demonstrated that the NIPAM gel dosimeter with the 5-3-3-10 formulation is suitable for verifying the dosimetry dose of proton beams.


Subject(s)
Proton Therapy , Radiation Dosimeters , Feasibility Studies , Protons , Acrylamides , Gels , Polymers