Med Eng Phys ; 35(4): 532-6, 2013 Apr.
Article in English | MEDLINE | ID: mdl-22998896

ABSTRACT

A concept for a partially implantable hearing device, for which the power supply and signal transmission are provided by an optical transmission path, is evaluated. The actuator is designed to fit into the round-window niche and to couple directly to the round-window membrane. Implantable hearing aids can be a suitable solution in the case of severe hearing loss, where conventional hearing aids often fail. However, the surgical effort for an implantation is comparatively high. Therefore, the objective of our work was to provide a hearing system which combines reliable coupling to the auditory system with an easy implantation technique. The actuator was designed as a piezoelectric thin-film cantilever. The optical transmission path was realised using an infrared light-emitting diode combined with an active receiver circuit. For a voltage of 1V, the unloaded actuator presents displacement amplitudes of 1µm up to a stimulus frequency of 25kHz and forces up to 0.2mN. Proportionally larger forces can be achieved by stacking single actuators. The overall transmission loss from the electrical input of the light-emitting diode driver to the mechanical output of the unloaded actuator was less than 25dB at 1kHz and maximum output.


Subject(s)
Ear Canal , Ossicular Prosthesis , Prosthesis Design , Infrared Rays
3.
Article in English | MEDLINE | ID: mdl-23221226

ABSTRACT

Bimorph structures are a standard method for transforming the high force of piezoelectric materials into a large deflection. In micro electromechanical systems (MEMS) applications, it is preferable to use structures consisting of a passive substrate (usually silicon) and one or more piezoelectric layers on the top. Such structures are called heterogeneous bimorphs or enakemesomorphs. In some MEMS applications- for example, for use as acoustic transducers-it is desirable to arrange such heterogeneous bimorphs in a circular shape, which results in trapezoidal cantilever structures. In this paper, an analytic dynamic description of such actuators is obtained. The resulting model is proved to be compatible with existing models for heterogeneous bimorphs with constant width. A comparison to a finite element analysis model of an exemplary layout shows divergences wholly within the same range as found for published models for constant-width structures.

SELECTION OF CITATIONS
SEARCH DETAIL
...