1.
Urology ; 154: 115-119, 2021 08.
Article in English | MEDLINE | ID: mdl-33774042

ABSTRACT

OBJECTIVE: To retrospectively investigate the safety of magnetic resonance imaging (MRI) in patients under sacral neuromodulation (InterStim II). METHODS: Data of patients who received a sacral neuromodulator at the urological department of a Swiss center of tertiary care from 2007 to 2018 and subsequently received at least 1 MRI with implanted device were retrospectively analyzed. Patient characteristics, data on implantation, MRI characteristics and complications potentially related to the MRI were analyzed. In addition, patient interviews were performed to verify the data gathered from patient records. RESULTS: A total of 55 consecutive patients with a median age of 48 years (range 16 - 80 years) and a total of 191 MRIs (median 3, range 1 - 13) were included to the study. The majority of MRIs (92%) were performed with 1.5 Tesla. The majority of the 1.5 Tesla (58%) as well as 3 Tesla (56%) MRIs assessed body regions other than the head. Complication possibly related to the MRI were only found in 2 (1%) MRI scans in two patients who reported on transient electrifying pain and heat sensation at the implantation site of the neuromodulator during MRI. CONCLUSION: MRI scans in patients with an implanted InterStim II sacral neuromodulator and with the device being turned off seem to be safe, even if they involve body regions other than the head, at least with 1.5 Tesla.


Subject(s)
Implantable Neurostimulators , Magnetic Resonance Imaging , Adolescent , Adult , Aged , Aged, 80 and over , Female , Humans , Male , Middle Aged , Retrospective Studies , Urinary Bladder Diseases/therapy , Young Adult
SELECTION OF CITATIONS
SEARCH DETAIL
...