1.
Klin Lab Diagn ; (12): 31-3, 2009 Dec.
Article in Russian | MEDLINE | ID: mdl-20141002

ABSTRACT

The Clauss method was recommended by the WHO to measure plasma fibrinogen levels. The purpose of this study was to assess the application of test systems made by the RENAM Research-and-Production Association (Moscow) on different types of clinical laboratory analyzers to measure fibrinogen by the Clauss method. The calibration line is shown to be of major value in obtaining correct results on each coagulometer. The findings suggest the accuracy and reproducibility of the determination of fibrinogen concentrations by the Clauss when the REHAM test systems are used to measure fibrinogen concentrations on various analyzers if accurate calibration lines are obtained.


Subject(s)
Fibrinogen/analysis , Blood Coagulation Tests , Calibration , Humans , Reagent Kits, Diagnostic
2.
Klin Lab Diagn ; (8): 29-30, 2006 Aug.
Article in Russian | MEDLINE | ID: mdl-17087244

ABSTRACT

A standard assay procedure for estimating antithrombin III activity was derived from the Abildgaard technique as follows: 1) to decrease errors at the preanalytical stage, by using preliminarily defibrinated lyophilized plasma-calibrator and 2) to increase measurement accuracy, by using the concentration of a fibrinogen preparation and simplifying the analytical procedure. The estimation of plasma antithrombin activity in patients with pulmonary thromboembolism or sepsis points to the changes in antithrombin III activity.


Subject(s)
Antithrombin III/analysis , Blood Coagulation Tests/standards , Humans , Reference Standards
3.
Klin Lab Diagn ; (7): 43-5, 2003 Jul.
Article in Russian | MEDLINE | ID: mdl-12934334

ABSTRACT

A method was elaborated enabling an isolation of a stable and highly active thrombin to be used in determining the thrombin time and fibrinogen concentration in plasma. The stability of the thus obtained preparations was studied, which showed no changes in the activity of their liquid variations for as long as 28 days at 2-8 degrees C.


Subject(s)
Fibrinogen/analysis , Thrombin Time , Thrombin/isolation & purification , Drug Stability , Drug Storage , Humans , Indicators and Reagents , Temperature , Thrombin/chemistry