Retina ; 25(5): 581-6, 2005.
Article in English | MEDLINE | ID: mdl-16077354

ABSTRACT

PURPOSE: To measure the amount and duration of indentation depth achieved with biodegradable poly-L/D-lactide 96/4 (PLA96) and silicone sponge implants. METHODS: Thirty rabbits underwent a scleral buckling procedure. A PLA96 buckling implant was used in 15 rabbits and a silicone sponge buckling implant was in 15 rabbits. A circumferential scleral buckling implant was sutured episclerally on the left eye of each rabbit, just temporal to the superior rectus muscle and 7 mm posterior to the limbus. Computed tomography was performed at 1 week, 3 months, and 5 months after surgery. RESULTS: The PLA96 buckling implant (implant diameter, 3-3.5 mm) used in this study created lower indentation than the silicone sponge implant (implant diameter, 4 mm). The indentation created by the PLA96 implant decreased over time compared with that created by the silicone implant. There were no complications related to either kind of implant. CONCLUSION: Both the silicone sponge implant and the PLA96 implant caused indentation that decreased in a comparable manner over the follow-up period (5 months).


Subject(s)
Absorbable Implants , Polyesters , Scleral Buckling/instrumentation , Silicone Elastomers , Animals , Male , Ocular Physiological Phenomena , Orbit/diagnostic imaging , Rabbits , Retinal Detachment/surgery , Sclera/surgery , Surgical Sponges , Suture Techniques , Time Factors , Tomography, X-Ray Computed