Radiat Prot Dosimetry ; 110(1-4): 619-22, 2004.
Article in English | MEDLINE | ID: mdl-15353718

ABSTRACT

We have developed a tiny neutron probe detector as a monitor of a thermal neutron flux for boron neutron capture therapy. The detector consists of an optical fibre and a small neutron probe. We have used a film-like ZnS(Ag) scintillator and a 6LiF neutron converter for the neutron probe. In order to improve the gamma-neutron discrimination ability, vacuum evaporation of 6LiF onto the ZnS(Ag) film has been done. In order to improve the neutron detection efficiency, we made use of a wavelength-shifting fibre as the probe material. The characteristics of the above two types of fibre probe detector have been evaluated experimentally.


Subject(s)
Boron Neutron Capture Therapy/instrumentation , Fiber Optic Technology/instrumentation , Neutrons/therapeutic use , Radiation Protection/instrumentation , Radiotherapy Planning, Computer-Assisted/instrumentation , Thermoluminescent Dosimetry/instrumentation , Transducers , Body Burden , Boron Neutron Capture Therapy/methods , Equipment Design , Equipment Failure Analysis , Fiber Optic Technology/methods , Gamma Rays , Humans , Miniaturization , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods , Relative Biological Effectiveness , Reproducibility of Results , Risk Assessment/methods , Sensitivity and Specificity , Thermoluminescent Dosimetry/methods