1.
J Appl Clin Med Phys ; 20(2): 114-120, 2019 Feb.
Article in English | MEDLINE | ID: mdl-30673145

ABSTRACT

The purpose of this study was to propose a verification method and results of intensity-modulated proton therapy (IMPT), using a commercially available heterogeneous phantom. We used a simple simulated head and neck and prostate phantom. An ionization chamber and radiochromic film were used for measurements of absolute dose and relative dose distribution. The measured doses were compared with calculated doses using a treatment planning system. We defined the uncertainty of the measurement point of the ionization chamber due to the effective point of the chamber and mechanical setup error as 2 mm and estimated the dose variation base on a 2 mm error. We prepared a HU-relative stopping power conversion table and fluence correction factor that were specific to the heterogeneous phantom. The fluence correction factor was determined as a function of depth and was obtained from the ratio of the doses in water and in the phantom at the same effective depths. In the simulated prostate plan, composite doses of measurements and calculations agreed within ±1.3% and the maximum local dose differences of each field were 10.0%. Composite doses in the simulated head and neck plan agreed within 4.0% and the maximum local dose difference for each field was 12.0%. The dose difference for each field came within 2% when taking the measurement uncertainty into consideration. In the composite plan, the maximum dose uncertainty was estimated as 4.0% in the simulated prostate plan and 5.8% in the simulated head and neck plan. Film measurements showed good agreement, with more than 92.5% of points passing a gamma value (3%/3 mm). From these results, the heterogeneous phantom should be useful for verification of IMPT by using a phantom-specific HU-relative stopping power conversion, fluence correction factor, and dose error estimation due to the effective point of the chamber.


Subject(s)
Neoplasms/radiotherapy , Phantoms, Imaging , Proton Therapy , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Humans , Organs at Risk/radiation effects , Radiotherapy Dosage
2.
J Appl Clin Med Phys ; 20(1): 258-264, 2019 Jan.
Article in English | MEDLINE | ID: mdl-30597762

ABSTRACT

The purpose of this study was to provide periodic quality assurance (QA) methods for respiratory-gated proton beam with a range modulation wheel (RMW) and to clarify the characteristics and long-term stability of the respiratory-gated proton beam. A two-dimensional detector array and a solid water phantom were used to measure absolute dose, spread-out Bragg peak (SOBP) width and proton range for monthly QA. SOBP width and proton range were measured using an oblique incidence beam to the lateral side of a solid water phantom and compared between with and without a gating proton beam. To measure the delay time of beam-on/off for annual QA, we collected the beam-on/off signals and the dose monitor-detected pulse. We analyzed the results of monthly QA over a 15-month period and investigated the delay time by machine signal analysis. The dose deviations at proximal, SOBP center and distal points were -0.083 ± 0.25%, 0.026 ± 0.20%, and -0.083 ± 0.35%, respectively. The maximum dose deviation between with and without respiratory gating was -0.95% at the distal point and other deviations were within ±0.5%. Proximal and SOBP center doses showed the same trend over a 15-month period. Delay times of beam-on/off for 200 MeV/SOBP 16 cm were 140.5 ± 0.8 ms and 22.3 ± 13.0 ms, respectively. Delay times for 160 MeV/SOBP 10 cm were 167.5 ± 15.1 ms and 19.1 ± 9.8 ms. Our beam delivery system with the RMW showed sufficient stability for respiratory-gated proton therapy and the system did not show dependency on the energy and the respiratory wave form. The delay times of beam-on/off were within expectations. The proposed QA methods will be useful for managing the quality of respiratory-gated proton beams and other beam delivery systems.


Subject(s)
Neoplasms/radiotherapy , Phantoms, Imaging , Proton Therapy/methods , Quality Assurance, Health Care/standards , Radiotherapy Planning, Computer-Assisted/methods , Respiratory-Gated Imaging Techniques/methods , Humans , Monte Carlo Method , Radiotherapy Dosage , Scattering, Radiation
3.
Pract Radiat Oncol ; 9(2): e149-e155, 2019 Mar.
Article in English | MEDLINE | ID: mdl-30336269

ABSTRACT

PURPOSE: Suppression of respiratory movement of the liver would be desirable for high-precision radiation therapy for liver tumors. We aimed to investigate the effect of our original device-free compressed shell fixation method and breathing instruction on suppression of respiratory movement. The characteristics of liver motion based on the movement of a fiducial marker were also analyzed. METHODS AND MATERIALS: First, respiratory amplitudes of the liver with the device-free compressed shell were analyzed from the data of 146 patients. The effect of this shell fixing method on liver movement was evaluated. Second, as another cohort study with 166 patients, interfractional internal motion of the liver for patients fixed in the shell was calculated using the fiducial marker coordinate data of images for position setting before daily irradiation. Third, in another 12 patients, intrafractional internal motion was calculated from the fiducial marker coordinate data using x-ray images before and after irradiation. RESULTS: The median respiratory movement without the shell, after fixing with the shell, and after instructing on the breathing method with the shell was 14.2 (interquartile range, 10.7-19.8), 11.5 (8.6-17.5), and 10.4 mm (7.3-15.8), respectively. Systematic and random errors of interfractional internal motion were all ≤2 mm in the left-right and anteroposterior directions and 3.7 and 3.0 mm, respectively, in the craniocaudal direction. Systematic and random errors of intrafractional internal motion were all ≤1.3 mm in the left-right and anteroposterior directions and 0.8 and 2.4 mm, respectively, in the craniocaudal direction. CONCLUSIONS: The device-free compressed shell fixation method was effective in suppressing the respiratory movement of the liver. Irradiation position matching using the fiducial marker can correct the interfractional internal motion on each day, which would contribute to the reduction of the margin to be given around the target.


Subject(s)
Liver Neoplasms/radiotherapy , Organ Motion , Patient Positioning/methods , Proton Therapy/methods , Radiotherapy, Image-Guided/methods , Fiducial Markers , Humans , Liver/diagnostic imaging , Liver/physiology , Liver Neoplasms/diagnostic imaging , Male , Prospective Studies , Radiotherapy Planning, Computer-Assisted/methods , Respiratory Mechanics/physiology , Supine Position , Tomography, X-Ray Computed