ABSTRACT
OBJECTIVE: To determine the concurrent validity and reliability of a low-cost spatiotemporal gait analysis system for clinical use in rehabilitation medicine. DESIGN: Cross-sectional study. SUBJECTS: Thirty-three healthy adults. METHODS: The spatiotemporal gait analysis system consists of a video camera placed perpendicular to a 10-m walkway and calibrated for spatial reference. The conditions evaluated in this study were: barefoot walking at comfortable and slow speed, toe and shod walking using a stationary camera setup and barefoot walking at comfortable speed using a moving camera setup. The GAITRite® was used as reference. RESULTS: High intraclass correlation coefficients (ICC≥ 0.97; 95% lower limit confidence intervals (95% CIs) ≥ 0.77) were found between systems for step and stride length, and step, stance and stride time, across setups and conditions. Standard error of measurement and Bland-Altman repeatability coefficients were ≤ 2.4% and ≤ 6.3%, respectively. A minimum of 4 footsteps was required to obtain ICC >0.90 and coefficient of variation < 10%. For double support and swing time, ICCs were generally low (ICC≥ 0.21). Inter-rater reliability was excellent for step length, step and stance time (ICC≥ 0.94; lower limit 95% CIs ≥ 0.86). CONCLUSION: The spatiotemporal gait analysis system is valid and reliable for assessing spatiotemporal parameters in different walking conditions. However, the validity of double support and swing time could not be confirmed.