1.
BJA Open ; 10: 100290, 2024 Jun.
Article in English | MEDLINE | ID: mdl-38966590

ABSTRACT

Introduction: Leakage of orogastric secretions past the cuff of a tracheal tube is a contributory factor in ventilator-associated pneumonia. Current bench test methods specified in the International Standard for Anaesthetic and Respiratory Equipment (EN ISO 5361:2023) to test cuff leakage involve using a glass or plastic rigid cylinder model of the trachea. There is a need for more realistic models to inform cuff leakage. Methods: We used human computerised tomography data and additive manufacturing (3D printing), combined with casting techniques to fabricate a bio-inspired synthetic tracheal model with analogous tissue characteristics. We conducted cuff leakage tests according to EN ISO 5361:2023 and compared results for high-volume low-pressure polyvinyl chloride and polyurethane cuffs between the rigid cylinder trachea with our bio-inspired model. Results: The tracheal model demonstrated close agreement with published tracheal tissue hardness for cartilaginous and membranous soft tissues. For high-volume low-pressure polyvinyl chloride cuffs the leakage rate was >50% lower in the bio-inspired tracheal model compared with the rigid cylinder model (151 [8] vs 261 [11] ml h-1). For high-volume low-pressure polyurethane cuffs, much lower leakage rates were observed than polyvinyl chloride cuffs in both models with leakage rates higher for the bio-inspired trachea model (0.1 [0.2] vs 0 [0] ml h-1). Conclusion: A reproducible tracheal model that incorporates the mechanical properties of the human trachea can be manufactured from segmented CT images and additive manufactured moulds, providing a useful tool to inform future cuff development, leakage testing for industrial applications, and clinical decision-making. There are differences between cuff leakage rates between the bio-inspired model and the rigid cylinder recommended in EN ISO 5361:2023. The bio-inspired model could lead to more accurate and realistic cuff leakage rate testing which would support manufacturers in refining their designs. Clinicians would then be able to choose better tracheal tubes based on the outcomes of this testing.

2.
Biomed Opt Express ; 13(1): 48-64, 2022 Jan 01.
Article in English | MEDLINE | ID: mdl-35154853

ABSTRACT

Incorrect endotracheal tube (ETT) cuff inflation pressure causes significant problems for intubated patients. The technical development and first in vivo use of a smart ETT for measurements at the cuff-trachea interface during mechanical ventilation are described. The intra-tracheal multiplexed sensing (iTraXS) ETT contains integrated optical fibre sensors to measure contact pressure and blood perfusion. The device is tested during mechanical ventilation in a porcine model (N=6). For contact pressure, signals were obtained in all 30 measurements. For perfusion, data could be obtained in all 33 measurements. In the 3 cases where the cuff was inflated to an artificially high-level, blood occlusion is observed.

3.
Sensors (Basel) ; 21(18)2021 Sep 15.
Article in English | MEDLINE | ID: mdl-34577383

ABSTRACT

Ammonia gas sensors were fabricated via layer-by-layer (LbL) deposition of diazo resin (DAR) and a binary mixture of tetrakis(4-sulfophenyl)porphine (TSPP) and poly(styrene sulfonate) (PSS) onto the core of a multimode U-bent optical fiber. The penetration of light transferred into the evanescent field was enhanced by stripping the polymer cladding and coating the fiber core. The electrostatic interaction between the diazonium ion in DAR and the sulfonate residues in TSPP and PSS was converted into covalent bonds using UV irradiation. The photoreaction between the layers was confirmed by UV-vis and Fourier transform infrared spectroscopy. The sensitivity of the optical fiber sensors to ammonia was linear when exposed to ammonia gases generated from aqueous ammonia solutions at a concentration of approximately 17 parts per million (ppm). This linearity extended up to 50 ppm when the exposure time (30 s) was shortened. The response and recovery times were reduced to 30 s with a 5-cycle DAR/TSPP+PSS (as a mixture of 1 mM TSPP and 0.025 wt% PSS in water) film sensor. The limit of detection (LOD) of the optimized sensor was estimated to be 0.31 ppm for ammonia in solution, corresponding to approximately 0.03 ppm of ammonia gas. It is hypothesized that the presence of the hydrophobic moiety of PSS in the matrix suppressed the effects of humidity on the sensor response. The sensor response was stable and reproducible over seven days. The PSS-containing U-bent fiber sensor also showed superior sensitivity to ammonia when examined alongside amine and non-amine analytes.


Subject(s)
Ammonia , Porphyrins , Humidity , Optical Fibers , Polystyrenes
4.
Sensors (Basel) ; 21(4)2021 Feb 18.
Article in English | MEDLINE | ID: mdl-33670602

ABSTRACT

A tip-based fibreoptic localised surface plasmon resonance (LSPR) sensor is reported for the sensing of volatile organic compounds (VOCs). The sensor is developed by coating the tip of a multi-mode optical fibre with gold nanoparticles (size: 40 nm) via a chemisorption process and further functionalisation with the HKUST-1 metal-organic framework (MOF) via a layer-by-layer process. Sensors coated with different cycles of MOFs (40, 80 and 120) corresponding to different crystallisation processes are reported. There is no measurable response to all tested volatile organic compounds (acetone, ethanol and methanol) in the sensor with 40 coating cycles. However, sensors with 80 and 120 coating cycles show a significant redshift of resonance wavelength (up to ~9 nm) to all tested volatile organic compounds as a result of an increase in the local refractive index induced by VOC capture into the HKUST-1 thin film. Sensors gradually saturate as VOC concentration increases (up to 3.41%, 4.30% and 6.18% in acetone, ethanol and methanol measurement, respectively) and show a fully reversible response when the concentration decreases. The sensor with the thickest film exhibits slightly higher sensitivity than the sensor with a thinner film. The sensitivity of the 120-cycle-coated MOF sensor is 13.7 nm/% (R2 = 0.951) with a limit of detection (LoD) of 0.005% in the measurement of acetone, 15.5 nm/% (R2 = 0.996) with an LoD of 0.003% in the measurement of ethanol and 6.7 nm/% (R2 = 0.998) with an LoD of 0.011% in the measurement of methanol. The response and recovery times were calculated as 9.35 and 3.85 min for acetone; 5.35 and 2.12 min for ethanol; and 2.39 and 1.44 min for methanol. The humidity and temperature crosstalk of 120-cycle-coated MOF was measured as 0.5 ± 0.2 nm and 0.5 ± 0.1 nm in the humidity range of 50-75% relative humidity (RH) and temperature range of 20-25 °C, respectively.

5.
Mikrochim Acta ; 187(12): 663, 2020 11 17.
Article in English | MEDLINE | ID: mdl-33201381

ABSTRACT

Highly sensitive and selective long-period grating (LPG) fiber-optic sensors modified with molecularly imprinted TiO2 nanothin films were fabricated. The films were deposited onto the surface of the optical fiber via liquid-phase deposition (LPD), using tetrakis(N-methylpyridinium-4-yl)porphyrin (TMPyP) as a template. Three LPG resonance bands were monitored during film deposition, which was of duration 4.5 h. Prior to template removal, heat treatment at 60 °C under high-humidity conditions led to an increase in refractive index of the TiO2 film, evidenced by changes in the central wavelengths of the attenuation bands. After template removal using HCl solution (0.01 M), the TMPyP-imprinted film-modified LPG sensor showed higher sensitivity to the template molecule than to structurally related guest molecules. This was measured at the 1st and 2nd resonance bands, with wavelengths ranging from 690 to 738 nm and 815 to 905 nm, respectively. No selective binding of the template was observed with a non-imprinted TiO2 film prepared in the same manner. Furthermore, the heat-treated imprinted films exhibited a substantial enhancement of photocatalytic activity for template irradiation. In particular, the self-cleaning property of the imprinted film-modified LPG sensor under ultraviolet irradiation led to highly efficient and selective binding to the template. The mechanism of the interaction between the template and the TiO2 matrix was investigated by UV-vis and Fourier-transform infrared (FTIR) spectroscopies. Additionally, morphological studies using scanning electron microscopy (SEM) were conducted. Graphical abstract.


Subject(s)
Fiber Optic Technology/instrumentation , Molecular Imprinting , Nanostructures/chemistry , Photochemical Processes , Titanium/chemistry , Catalysis , Microscopy, Electron, Scanning , Spectroscopy, Fourier Transform Infrared , Spectrum Analysis/methods
6.
Analyst ; 145(13): 4504-4511, 2020 Jul 07.
Article in English | MEDLINE | ID: mdl-32409797

ABSTRACT

Dabrafenib is one of the most widely used of the new generation of targeted anti-cancer drugs. However, its therapeutic window varies for different patients and so there is an unmet need for methods to monitor the dose of drug which the patient receives and at the specific site where it acts. In the case of cancers, it is critical to measure the concentration of drug not just in the bloodstream overall, but in or near tumours, as these will not be the same over multiple time periods. A novel sensor based on an optical fibre long period grating (LPG) modified with a molecular imprinted polymer (MIP) has been developed with the ultimate aim of achieving minimally invasive measurements of Dabrafenib at the tumour site. A molecularly imprinted polymer specific for Dabrafenib was coated on a methacryloylalkoxysilane-functionalised optical fibre long period grating. In vitro experimental results demonstrate that the Dabrafenib sensitivity is 15.2 pm (µg mL-1)-1 (R2 = 0.993) with a limit of detection (LoD) of 74.4 µg mL-1 in serum solution. Moreover, the proposed sensor shows selective response to Dabrafenib over structurally similar 2-Aminoquinoline.


Subject(s)
Antineoplastic Agents/blood , Imidazoles/blood , Molecularly Imprinted Polymers/chemistry , Optical Fibers , Oximes/blood , Animals , Cattle , Limit of Detection , Spectrophotometry/instrumentation , Spectrophotometry/methods
7.
Sensors (Basel) ; 19(10)2019 May 17.
Article in English | MEDLINE | ID: mdl-31109017

ABSTRACT

The development of reliable, affordable and efficient sensors is a key step in providing tools for efficient monitoring of critical environmental parameters. This review focuses on the use of tapered optical fibres as an environmental sensing platform. Tapered fibres allow access to the evanescent wave of the propagating mode, which can be exploited to facilitate chemical sensing by spectroscopic evaluation of the medium surrounding the optical fibre, by measurement of the refractive index of the medium, or by coupling to other waveguides formed of chemically sensitive materials. In addition, the reduced diameter of the tapered section of the optical fibre can offer benefits when measuring physical parameters such as strain and temperature. A review of the basic sensing platforms implemented using tapered optical fibres and their application for development of fibre-optic physical, chemical and bio-sensors is presented.

8.
Sensors (Basel) ; 17(2)2017 Feb 08.
Article in English | MEDLINE | ID: mdl-28208691

ABSTRACT

A long period grating (LPG) modified with a mesoporous film infused with a calixarene as a functional compound was employed for the detection of individual volatile organic compounds (VOCs) and their mixtures. The mesoporous film consisted of an inorganic part, SiO2 nanoparticles (NPs), along with an organic moiety of poly(allylamine hydrochloride) polycation PAH, which was finally infused with the functional compound, p-sulphanato calix[4]arene (CA[4]) or p-sulphanato calix[8]arene (CA[8]). The LPG sensor was designed to operate at the phase matching turning point to provide the highest sensitivity. The sensing mechanism is based on the measurement of the refractive index (RI) change induced by a complex of the VOCs with calixarene. The LPG, modified with a coating of 5 cycles of (SiO2 NPs/PAH) and infused with CA[4] or CA[8], was exposed to chloroform, benzene, toluene and acetone vapours. The British Standards test of the VOCs emissions from material (BS EN ISO 16000-9:2006) was used to test the LPG sensor performance.

9.
IEEE Trans Biomed Eng ; 63(9): 1985-1992, 2016 09.
Article in English | MEDLINE | ID: mdl-26829771

ABSTRACT

OBJECTIVE: To develop a compact probe that can be used to monitor humidity in ventilator care equipment. A mesoporous film of alternate layers of Poly(allylamine hydrochloride) (PAH) and silica (SiO2) nanoparticles (bilayers), deposited onto an optical fibre was used. The sensing film behaves as a Fabry-Perot cavity of low-finesse where the absorption of water vapour changes the optical thickness and produces a change in reflection proportional to humidity. METHODS: The mesoporous film was deposited upon the cleaved tip of an optical fibre using the layer-by-layer method. The sensor was calibrated in a bench model against a commercially available capacitive sensor. The sensitivity and response time were assessed in the range from 5 % relative humidity (RH) to 95%RH for different numbers of bilayers up to a maximum of nine. RESULTS: The sensitivity increases with the number of bilayers deposited; sensitivity of 2.28 mV/%RH was obtained for nine bilayers. The time constant of the response was 1.13 s ± 0.30 s which is faster than the commercial device (measured as 158 s). After calibration, the optical fibre humidity sensor was utilised in a bench top study employing a mechanical ventilator. The fast response time enabled changes in humidity in individual breaths to be resolved. CONCLUSION: Optical fibre sensors have the potential to be used to monitor breath to breath humidity during ventilator care. SIGNIFICANCE: Control of humidity is an essential part of critical respiratory care and the developed sensor provides a sensitive, compact and fast method of humidity monitoring.


Subject(s)
Fiber Optic Technology/instrumentation , Humidity , Nanoconjugates/chemistry , Refractometry/instrumentation , Respiration, Artificial/instrumentation , Transducers , Equipment Design , Equipment Failure Analysis , Interferometry/instrumentation , Membranes, Artificial , Nanoconjugates/ultrastructure , Optical Fibers , Polyamines/chemistry , Reproducibility of Results , Sensitivity and Specificity , Silicon Dioxide/chemistry
10.
Analyst ; 139(9): 2229-36, 2014 May 07.
Article in English | MEDLINE | ID: mdl-24634909

ABSTRACT

An optical fibre long period grating (LPG) sensor modified with molecularly imprinted polymer nanoparticles (nanoMIPs) for the specific detection of antibiotics is presented. The operation of the sensor is based on the measurement of changes in refractive index induced by the interaction of nanoMIPs deposited onto the cladding of the LPG with free vancomycin (VA). The binding of nanoMIPs to vancomycin was characterised by a binding constant of 4.3 ± 0.1 × 10(-8) M. The lowest concentration of analyte measured by the fibre sensor was 10 nM. In addition, the sensor exhibited selectivity, as much smaller responses were obtained for high concentrations (∼700 µM) of other commonly prescribed antibiotics such as amoxicillin, bleomycin and gentamicin. In addition, the response of the sensor was characterised in a complex matrix, porcine plasma, spiked with 10 µM of VA.


Subject(s)
Anti-Bacterial Agents/analysis , Molecular Imprinting , Nanoparticles , Polymers/chemistry , Vancomycin/analysis , Animals , Anti-Bacterial Agents/blood , Limit of Detection , Swine , Vancomycin/blood
11.
Anal Chim Acta ; 694(1-2): 142-50, 2011 May 23.
Article in English | MEDLINE | ID: mdl-21565315

ABSTRACT

TiO(2) nano-thin films with imprinted (R)- and (S)-enantiomers of propranolol, 1,1'-bi-naphthol, and 2-(4-isobutylphenyl)-propionic acid were fabricated on quartz plates by spin-coating their solutions with Ti(O-(n)Bu)(4) in a toluene-ethanol mixture (1:1, v/v). After template removal, the imprinted films showed better binding for original templates than to the corresponding enantiomers. The assessment of template incorporation, template removal, and re-binding was conducted through UV-vis measurements. Significant enhancement of enantioselectivity was achieved by optimization of the film thickness and by heat-treatment of the imprinted films. After subtraction of non-specific binding, the optimized films provided chiral recognition with the enantioselectivity of almost 100% for (R)-propranolol and 95% for (S)-propranolol.


Subject(s)
Molecular Imprinting/methods , Nanostructures/chemistry , Titanium/chemistry , Ibuprofen , Naphthols/chemistry , Propionates/chemistry , Propranolol/chemistry , Spectrophotometry, Ultraviolet/methods , Stereoisomerism
12.
Opt Express ; 18(12): 13227-38, 2010 Jun 07.
Article in English | MEDLINE | ID: mdl-20588452

ABSTRACT

A novel approach to chemical application of long period grating (LPG) optical fibers was demonstrated, which were modified with a film nanoassembled by the alternate deposition of SiO(2) nanoparticles (SiO(2) NPs) and poly(diallyldimethyl ammonium chloride) (PDDA). Nanopores of the sensor film could be used for sensitive adsorption of chemical species in water, which induced the changes in the refractive index (RI) of the light propagating in the cladding mode of the optical fiber, with a concomitant effect on the transmission spectrum in the LPG region. The prepared fiber sensor was highly sensitive to the change in the RI of the surrounding medium and the response time was very fast within 10 s. In addition, chemical infusion into the film was tested using a porphyrin compound, tetrakis-(4-sulfophenyl)porphine (TSPP), which could be saturated within a few min. The lowest detectable concentration of the TSPP analyte was 10 microM. The TSPP infusion led to the development of well-pronounced dual resonance bands, indicating a large increase in the optical thickness of the film. The RI of the film was dramatically increased from 1.200 to ca. 1.540.

13.
Anal Chem ; 82(6): 2228-36, 2010 Mar 15.
Article in English | MEDLINE | ID: mdl-20170084

ABSTRACT

Quartz crystal microbalance (QCM) gas sensors based on the alternate adsorption of TiO(2) and polyacrilic acid (PAA) were developed for the sensitive detection of amine odors. Individual TiO(2) gel layers could be regularly assembled with a thickness of approximately 0.3 nm by the gas-phase surface sol-gel process (GSSG). The thickness of the poly(acrylic acid) (PAA) layer is dependent on its molecular weight, showing different thicknesses of approximately 0.4 nm for PAA(25) (Mw 250,000) and 0.6-0.8 nm for PAA(400) (Mw 4,000,000). The QCM sensors showed a linear response to ammonia in the concentration range 0.3-15 ppm, depending on the deposition cycle of the alternate TiO(2)/PAA layer. The ammonia binding is based on the acid-base interaction to the free carboxylic acid groups of PAA and the limit of detection (LOD) of the 20-cycle TiO(2)/PAA(400) film was estimated to be 0.1 ppm when exposed to ammonia. The sensor response was very fast and stable in a wide relative humidity (rH) range of 30-70%, showing almost the same frequency changes at a given concentration of ammonia. Sensitivity to n-butylamine and ammonia was higher than to pyridine, which is owing to the difference of molecular weight and basicity of the amine analytes. The alternate TiO(2)/PAA(400) films have a highly effective ability to capture amine odors, and the ambient ammonia concentration of 15 ppm could be condensed up to approximately 20,000 ppm inside the films.