J Synchrotron Radiat ; 14(Pt 1): 151-7, 2007 Jan.
Article in English | MEDLINE | ID: mdl-17211083

ABSTRACT

Results obtained using a hybrid pixel photon-counting detector in powder diffraction experiments are presented. The detector works at room temperature and its dynamic response ranges from 0.01 photons pixel(-1) s(-1) up to 10(6) photons pixel(-1) s(-1). The pixel sizes are 0.33 mm x 0.33 mm for a total area of 68 mm x 68 mm. On recording high-resolution diffraction patterns of powders, a reduction of the experimental time by more than a factor of 20 is obtained without loss of data quality. The example of an X-zeolite shows that such detectors can be used for very demanding anomalous experiments. In situ experiments of quenching liquid oxides show that frames of 0.01 s can be achieved for studying such processes.


Subject(s)
Photons , Powder Diffraction/instrumentation , Radiometry/instrumentation , Signal Processing, Computer-Assisted/instrumentation , Transducers , Equipment Design , Equipment Failure Analysis , Pilot Projects , Powder Diffraction/methods , Radiation Dosage , Radiometry/methods , Semiconductors