Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Publication year range
1.
Radiol Phys Technol ; 16(2): 227-234, 2023 Jun.
Article in English | MEDLINE | ID: mdl-36947352

ABSTRACT

Treatment planning systems that use the Monte Carlo algorithm can calculate the dose to the medium (Dm) in non-water-equivalent tissues such as bones. However, Dm cannot be verified using actual measurements; therefore, it is necessary to develop tissue-equivalent dosimeters. In this study, we developed a bone-equivalent polymer gel dosimeter (BPGD) that can measure the dose absorbed by the bone and investigated its sensitivity. The BPGDs were prepared by adding 3.0 mol of calcium hydrogen phosphate dihydrate as a component of bone to an improved dose-sensitive polyacrylamide gelatin and tetrakis hydroxymethyl phosphonium chloride (iPAGAT). One day after preparation, the BPGDs were irradiated with a field size of 15 × 15 cm2 using a 10 MV X-ray beam to evaluate the dose sensitivity, dose-rate dependence, and dose-integration dependence. One day after dose exposure, the BPGDs were scanned using a 0.4 T MRI APERTO Eterna (Hitachi, Tokyo, Japan) to obtain R2 values. The difference between the R2 values of 6 Gy and 0 Gy was up to 5 s-1, and the R2 curve plateaued in the high-dose region. Moreover, the BPGD did not depend on the integration of the dose and dose rates. Therefore, the BPGDs that we developed can determine the radiation dose to bones.


Subject(s)
Algorithms , Radiation Dosimeters , Japan , Monte Carlo Method , Polymers , Radiometry , Gels
SELECTION OF CITATIONS
SEARCH DETAIL
...