Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Pak J Pharm Sci ; 23(4): 379-83, 2010 Oct.
Article in English | MEDLINE | ID: mdl-20884450

ABSTRACT

A simple, specific and accurate reverse phase liquid chromatographic method has been developed for the simultaneous determination of naproxen and ranitidine HCl. Both the drugs are official with British Pharmacopoeia 2007, but do not involve simultaneous determination of naproxen and ranitidine HCl. The separation was carried out using 4.6 × 250 mm Symmetry Shield TM RP 18 with a particle diameter of 5 µm and mobile phase containing 0.1M orthophosphoric acid: methanol (35:65, pH 3.1) in isocratic mode. The flow rate was 1.00 ml/min and effluent was monitored at 240 nm. The retention times (average) of ranitidine HCl and naproxen were 2.36 min and 12.39 min, respectively. The linearity for naproxen and ranitidine HCl was in the range of 5-35 µg/ml and 1.5-12 µg/ml, respectively. The potencies of naproxen and ranitidine HCl were found 99.40 % and 99.48 %, respectively. The proposed method was validated and successfully applied to the estimation of naproxen and ranitidine HCl in newly formulated combined tablet and in plasma.


Subject(s)
Anti-Inflammatory Agents, Non-Steroidal/analysis , Histamine H2 Antagonists/analysis , Naproxen/analysis , Ranitidine/analysis , Anti-Inflammatory Agents, Non-Steroidal/blood , Buffers , Calibration , Chemistry, Pharmaceutical , Chromatography, High Pressure Liquid , Drug Combinations , Histamine H2 Antagonists/blood , Humans , Naproxen/blood , Ranitidine/blood , Reference Standards , Reproducibility of Results , Spectrophotometry, Ultraviolet , Tablets
SELECTION OF CITATIONS
SEARCH DETAIL
...