ABSTRACT
The requirement of an independent verification of the monitor units (MU) or time calculated to deliver the prescribed dose to a patient has been a mainstay of radiation oncology quality assurance. The need for and value of such a verification was obvious when calculations were performed by hand using look-up tables, and the verification was achieved by a second person independently repeating the calculation. However, in a modern clinic using CT/MR/PET simulation, computerized 3D treatment planning, heterogeneity corrections, and complex calculation algorithms such as convolution/superposition and Monte Carlo, the purpose of and methodology for the MU verification have come into question. In addition, since the verification is often performed using a simpler geometrical model and calculation algorithm than the primary calculation, exact or almost exact agreement between the two can no longer be expected. Guidelines are needed to help the physicist set clinically reasonable action levels for agreement. This report addresses the following charges of the task group: (1) To re-evaluate the purpose and methods of the "independent second check" for monitor unit calculations for non-IMRT radiation treatment in light of the complexities of modern-day treatment planning. (2) To present recommendations on how to perform verification of monitor unit calculations in a modern clinic. (3) To provide recommendations on establishing action levels for agreement between primary calculations and verification, and to provide guidance in addressing discrepancies outside the action levels. These recommendations are to be used as guidelines only and shall not be interpreted as requirements.
Subject(s)
Physics , Radiotherapy/methods , Research Report , Societies , Algorithms , Computers , Humans , Quality Control , Radiometry , Radiosurgery , Radiotherapy Planning, Computer-Assisted , Reproducibility of Results , Research Design , United StatesABSTRACT
The concept of in-air output ratio (Sc) was introduced to characterize how the incident photon fluence per monitor unit (or unit time for a Co-60 unit) varies with collimator settings. However, there has been much confusion regarding the measurement technique to be used that has prevented the accurate and consistent determination of Sc. The main thrust of the report is to devise a theoretical and measurement formalism that ensures interinstitutional consistency of Sc. The in-air output ratio, Sc, is defined as the ratio of primary collision water kerma in free-space, Kp, per monitor unit between an arbitrary collimator setting and the reference collimator setting at the same location. Miniphantoms with sufficient lateral and longitudinal thicknesses to eliminate electron contamination and maintain transient electron equilibrium are recommended for the measurement of Sc. The authors present a correction formalism to extrapolate the correct Sc from the measured values using high-Z miniphantom. Miniphantoms made of high-Z material are used to measure Sc for small fields (e.g., IMRT or stereotactic radiosurgery). This report presents a review of the components of Sc, including headscatter, source-obscuring, and monitor-backscattering effects. A review of calculation methods (Monte Carlo and empirical) used to calculate Sc for arbitrary shaped fields is presented. The authors discussed the use of Sc in photon dose calculation algorithms, in particular, monitor unit calculation. Finally, a summary of Sc data (from RPC and other institutions) is included for QA purposes.