1.
Proc Inst Mech Eng H ; 235(12): 1463-1470, 2021 Dec.
Article in English | MEDLINE | ID: mdl-34278841

ABSTRACT

The aim of this experimental study was to assess the biomechanical performance of a novel C1 posterior arch (C1PA) clamp compared with C1 lateral mass (C1LM) screws in constructs used to treat atlantoaxial instability. These constructs had either C2 pedicle (C2P) screws or C2 translaminar (C2TL) screws. Eight fresh-frozen human cadaveric ligamentous spine specimens (C0-C3) were tested under six conditions: the intact state, the destabilized state after a simulated odontoid fracture, and when instrumented with four constructs (C1LM-C2P, C1LM-C2TL, C1PA-C2P, C1PA-C2TL). Each specimen was tested in a spinal loading simulator that separately applied axial rotation, flexion-extension and lateral bending. In each test condition, displacement controlled angular motion was applied in both directions at a speed of 2 deg/s until a resulting moment of 1.5 Nm was achieved. The measured ranges of motion (ROM) of the C1-C2 segments were compared for each test condition using nonparametric Friedman tests. The destabilized state had significantly more C1-C2 motion (p < 0.05) than the intact state in all cases, and all constructs greatly reduced this motion. C2 pedicle screw constructs that used the C1PA clamp had significantly less C1-C2 motion (p < 0.05) than those with C1LM screws in flexion-extension as well as axial rotation and no statistically significant difference was detected in lateral bending. C2 translaminar screw constructs that used the C1PA clamp had significantly less C1-C2 motion (p < 0.05) than those with C1LM screws in flexion-extension and no statistically significant difference was detected in axial rotation or in lateral bending. Data from the current study suggested that constructs using the novel C1PA clamp would provide as good, or improved, biomechanical stability to the C1-C2 segment compared with constructs using C1LM screws.


Subject(s)
Atlanto-Axial Joint , Spinal Fusion , Atlanto-Axial Joint/surgery , Biomechanical Phenomena , Cadaver , Cervical Vertebrae , Humans
2.
Spine J ; 17(11): 1755-1764, 2017 11.
Article in English | MEDLINE | ID: mdl-28673824

ABSTRACT

BACKGROUND CONTEXT: Predicting physiological range of motion (ROM) using a finite element (FE) model of the upper cervical spine requires the incorporation of ligament laxity. The effect of ligament laxity can be observed only on a macro level of joint motion and is lost once ligaments have been dissected and preconditioned for experimental testing. As a result, although ligament laxity values are recognized to exist, specific values are not directly available in the literature for use in FE models. PURPOSE: The purpose of the current study is to propose an optimization process that can be used to determine a set of ligament laxity values for upper cervical spine FE models. Furthermore, an FE model that includes ligament laxity is applied, and the resulting ROM values are compared with experimental data for physiological ROM, as well as experimental data for the increase in ROM when a Type II odontoid fracture is introduced. DESIGN/SETTING: The upper cervical spine FE model was adapted from a 50th percentile male full-body model developed with the Global Human Body Models Consortium (GHBMC). FE modeling was performed in LS-DYNA and LS-OPT (Livermore Software Technology Group) was used for ligament laxity optimization. METHODS: Ordinate-based curve matching was used to minimize the mean squared error (MSE) between computed load-rotation curves and experimental load-rotation curves under flexion, extension, and axial rotation with pure moment loads from 0 to 3.5 Nm. Lateral bending was excluded from the optimization because the upper cervical spine was considered to be primarily responsible for flexion, extension, and axial rotation. Based on recommendations from the literature, four varying inputs representing laxity in select ligaments were optimized to minimize the MSE. Funding was provided by the Natural Sciences and Engineering Research Council of Canada as well as GHMBC. The present study was funded by the Natural Sciences and Engineering Research Council of Canada to support the work of one graduate student. There are no conflicts of interest to be reported. RESULTS: The MSE was reduced to 0.28 in the FE model with optimized ligament laxity compared with an MSE 0f 4.16 in the FE model without laxity. In all load cases, incorporating ligament laxity improved the agreement between the ROM of the FE model and the ROM of the experimental data. The ROM for axial rotation and extension was within one standard deviation of the experimental data. The ROM for flexion and lateral bending was outside one standard deviation of the experimental data, but a compromise was required to use one set of ligament laxity values to achieve a best fit to all load cases. Atlanto-occipital motion was compared as a ratio to overall ROM, and only in extension did the inclusion of ligament laxity not improve the agreement. After a Type II odontoid fracture was incorporated into the model, the increase in ROM was consistent with experimental data from the literature. CONCLUSIONS: The optimization approach used in this study provided values for ligament laxities that, when incorporated into the FE model, generally improved the ROM response when compared with experimental data. Successfully modeling a Type II odontoid fracture showcased the robustness of the FE model, which can now be used in future biomechanics studies.


Subject(s)
Cervical Vertebrae/physiology , Computer Simulation , Ligaments/physiology , Range of Motion, Articular , Biomechanical Phenomena , Finite Element Analysis , Humans , Male , Rotation