1.
Biosens Bioelectron ; 92: 465-471, 2017 Jun 15.
Article in English | MEDLINE | ID: mdl-27856162

ABSTRACT

Quantitative microfluidic point-of-care testing has been translated into clinical applications to support a prompt decision on patient treatment. A nanointerstice-driven filling technique has been developed to realize the fast and robust filling of microfluidic channels with liquid samples, but it has failed to provide a consistent filling time owing to the wide variation in liquid viscosity, resulting in an increase in quantification errors. There is a strong demand for simple and quick flow control to ensure accurate quantification, without a serious increase in system complexity. A new control mechanism employing two-beam refraction and one solenoid valve was developed and found to successfully generate digitized filling flow, completely free from errors due to changes in viscosity. The validity of digitized filling flow was evaluated by the immunoassay, using liquids with a wide range of viscosity. This digitized microfluidic filling flow is a novel approach that could be applied in conventional microfluidic point-of-care testing.


Subject(s)
Microfluidic Analytical Techniques/instrumentation , Animals , Antibodies, Immobilized/chemistry , Biosensing Techniques/instrumentation , Equipment Design , Immunoassay/instrumentation , Immunoglobulin G/chemistry , Mice , Point-of-Care Testing , Viscosity