1.
Can J Anaesth ; 59(3): 288-94, 2012 Mar.
Article in English | MEDLINE | ID: mdl-22183297

ABSTRACT

PURPOSE: Several fluids are available for volume therapy to address hypovolemia. We focus on two hydroxyethyl starches (HES) available for volume expansion in Canada, HES 130/0.4 (Voluven®) and HES 260/0.45 (Pentaspan®). Although information is available regarding their pharmacokinetic and risk/benefit profiles, this paper examines their viscous properties. METHODS: Dynamic viscosities of HES 130/0.4 and HES 260/0.45 were measured through capillary viscometry at 21°C and 37°C. The viscosities of the solutions were then measured through a closed flow loop at room temperature across physiologically relevant flow rates that maintained a laminar flow regime. RESULTS: Measured dynamic viscosity through capillary viscometry for HES 130/0.4 and HES 260/0.45 was 2.76 centipoises (cP) and 7.62 cP, respectively, at 21°C decreasing to 1.74 cP and 4.25 cP, respectively, at 37°C. Pipe flow analysis found that HES 130/0.4 (expiry 02/13) and HES 260/0.45 (expiry 10/10) displayed marginal variation in viscosity suggesting Newtonian behaviour. However, a sample of HES 130/0.4 (expiry 10/10) displayed an appreciable increase in viscosity (13%) at higher flow rates suggesting shear thickening behaviour. CONCLUSION: This study represents an innovative characterization of not only the viscosity of two commonly utilized HES solutions but also their viscous behaviour across physiologically relevant flow rates. The shear thickening behaviour of a sample of HES 130/0.40 (expiry 10/10) at high flow rates was not expected, and the effect this result may have on endothelial cell function is unknown.


Subject(s)
Hydroxyethyl Starch Derivatives/chemistry , Plasma Substitutes/chemistry , Shear Strength , Viscosity
2.
Cardiovasc Eng ; 10(3): 128-35, 2010 Sep.
Article in English | MEDLINE | ID: mdl-20717726

ABSTRACT

Endovascular stents are commonly used to manage arterial diseases such as Aortic Abdominal Aneurysm (AAA), aortic dissection and coarctation. The radial force the stent applies to the vessel must be large enough to resist stent migration, but not so large that the mechanical stimulus initiates adverse vessel remodeling. We employed two approaches to characterize the radial force of Gianturco stents: first, by applying an external pressure to the stent and, second, by measuring the force exerted by the stent when deployed. From the second approach, we determined the force exerted at various area reductions that correspond to clinically relevant diameter oversizings. In this study, stent stiffness was determined from the force-area reduction curves. Comparing similar stents of various diameters revealed that smaller diameter stent had greater radial force and stiffness than larger diameter stents. Comparing similar stents of various lengths revealed that stents with longer lengths (and greater number of wires) has greater force and stiffness. Overlapping two stents increased the force and stiffness to values greater than the sum of those parameters for the individual stents. These data may have important clinical implications for understanding the effect of oversized and overlapped stents on vessel mechanics.


Subject(s)
Blood Vessel Prosthesis , Models, Theoretical , Stents , Computer Simulation , Computer-Aided Design , Elastic Modulus , Equipment Design , Equipment Failure Analysis , Stress, Mechanical