J Forensic Sci ; 54(2): 415-21, 2009 Mar.
Article in English | MEDLINE | ID: mdl-19187461

ABSTRACT

Operator exposure to backscatter radiation while using an Aribex NOMAD radiation emitting device (a portable, self-contained, cordless, hand-held dental X-ray unit) was determined while the operator employed various typical and atypical use scenarios during the exposure of 715 digital and/or film-based dental radiographs and 200 study control exposures. Study data was compared to the radiation safety occupational exposure annual Maximum Permissive Dose (MPD) of 50 mSv (5000 mrem) to determine the possible exposure risk to an unprotected operator using this device. The results showed the reproductive organs received the highest dose and the thyroid the least. The average operator whole body dose for the study was determined to be 0.047 mSv (4.47 mrem) or 0.09% of the annual MPD. Extrapolating the data as an expression of averaged annual operator exposure resulted in a whole body dose of 0.4536 mSv (45.36 mrem) or 0.9% of the annual MPD, These results are well below established occupation exposure limits and are compatible with those published by the manufacturer.


Subject(s)
Occupational Exposure/adverse effects , Radiography, Dental/instrumentation , Scattering, Radiation , Humans , Occupational Exposure/analysis , Radiation Dosage