Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 20 de 50
Filter
1.
Phys Med Biol ; 69(11)2024 May 17.
Article in English | MEDLINE | ID: mdl-38608644

ABSTRACT

Purpose. Radiation delivered over ultra-short timescales ('FLASH' radiotherapy) leads to a reduction in normal tissue toxicities for a range of tissues in the preclinical setting. Experiments have shown this reduction occurs for total delivery times less than a 'critical' time that varies by two orders of magnitude between brain (∼0.3 s) and skin (⪆10 s), and three orders of magnitude across different bowel experiments, from ∼0.01 to ⪆(1-10) s. Understanding the factors responsible for this broad variation may be important for translation of FLASH into the clinic and understanding the mechanisms behind FLASH.Methods.Assuming radiolytic oxygen depletion (ROD) to be the primary driver of FLASH effects, oxygen diffusion, consumption, and ROD were evaluated numerically for simulated tissues with pseudorandom vasculatures for a range of radiation delivery times, capillary densities, and oxygen consumption rates (OCR's). The resulting time-dependent oxygen partial pressure distribution histograms were used to estimate cell survival in these tissues using the linear quadratic model, modified to incorporate oxygen-enhancement ratio effects.Results. Independent of the capillary density, there was a substantial increase in predicted cell survival when the total delivery time was less than the capillary oxygen tension (mmHg) divided by the OCR (expressed in units of mmHg/s), setting the critical delivery time for FLASH in simulated tissues. Using literature OCR values for different normal tissues, the predicted range of critical delivery times agreed well with experimental values for skin and brain and, modifying our model to allow for fluctuating perfusion, bowel.Conclusions. The broad three-orders-of-magnitude variation in critical irradiation delivery times observed inin vivopreclinical experiments can be accounted for by the ROD hypothesis and differences in the OCR amongst simulated normal tissues. Characterization of these may help guide future experiments and open the door to optimized tissue-specific clinical protocols.


Subject(s)
Oxygen , Oxygen/metabolism , Kinetics , Time Factors , Radiotherapy/methods , Humans , Models, Biological , Oxygen Consumption/radiation effects , Cell Survival/radiation effects
2.
Article in English | MEDLINE | ID: mdl-38445180

ABSTRACT

Purpose: An integrated magnetic resonance scanner and linear accelerator (MR-linac) was implemented with daily online adaptive radiation therapy (ART). This study evaluated patient-reported experiences with their overall hospital care as well as treatment in the MR-linac environment. Methods: Patients pre-screened for MR eligibility and claustrophobia were referred to simulation on a 1.5 T MR-linac. Patient-reported experience measures were captured using two validated surveys. The 15-item MR-anxiety questionnaire (MR-AQ) was administered immediately after the first treatment to rate MR-related anxiety and relaxation. The 40-item satisfaction with cancer care questionnaire rating doctors, radiation therapists, the services and care organization and their outpatient experience was administered immediately after the last treatment using five-point Likert responses. Results were analyzed using descriptive statistics. Results: 205 patients were included in this analysis. Multiple sites were treated across the pelvis and abdomen with a median treatment time per fraction of 46 and 66 min respectively. Patients rated MR-related anxiety as "not at all" (87%), "somewhat" (11%), "moderately" (1%) and "very much so" (1%). Positive satisfaction responses ranged from 78 to 100% (median 93%) across all items. All radiation therapist-specific items were rated positively as 96-100%. The five lowest rated items (range 78-85%) were related to general provision of information, coordination, and communication. Overall hospital care was rated positively at 99%. Conclusion: In this large, single-institution prospective cohort, all patients had low MR-related anxiety and completed treatment as planned despite lengthy ART treatments with the MR-linac. Patients overall were highly satisfied with their cancer care involving ART using an MR-linac.

3.
Pract Radiat Oncol ; 2023 Nov 14.
Article in English | MEDLINE | ID: mdl-37967747

ABSTRACT

PURPOSE: This work aims at reviewing challenges and pitfalls in proton facility design related to equipment upgrade or replacement. Proton therapy was initially developed at research institutions in the 1950s which ushered in the use of hospital-based machines in 1990s. We are approaching an era where older commercial machines are reaching the end of their life and require replacement. The future widespread application of proton therapy depends on cost reduction; customized building design and installation are significant expenses. METHODS AND MATERIALS: We take this opportunity to discuss how commercial proton machines have been installed and how buildings housing the equipment have been designed. RESULTS: Data on dimensions and weights of the larger components of proton systems (cyclotron main magnet and gantries) are presented and innovative, non-gantry-based, patient positioning systems are discussed. CONCLUSIONS: We argue that careful consideration of the building design to include larger elevators, hoistways from above, wide corridors and access slopes to below grade installations, generic vault and treatment room layouts to accommodate multiple vendor's equipment, and modular system design can provide specific benefits during planning, installation, maintenance, and replacement phases of the project. Room temperature magnet coils can be constructed in a more modular manner: a potential configuration is presented. There is scope for constructing gantries and magnet yokes from smaller modular sub-units. These considerations would allow a hospital to replace a commercial machine at its end of life in a manner similar to a linac.

4.
Crit Rev Oncol Hematol ; 192: 104143, 2023 Dec.
Article in English | MEDLINE | ID: mdl-37742884

ABSTRACT

With increasing reliance on technology in oncology, the impact of digital clinical decision support (CDS) tools needs to be examined. A systematic review update was conducted and peer-reviewed literature from 2016 to 2022 were included if CDS tools were used for live decision making and comparatively assessed quantitative outcomes. 3369 studies were screened and 19 were included in this updated review. Combined with a previous review of 24 studies, a total of 43 studies were analyzed. Improvements in outcomes were observed in 42 studies, and 34 of these were of statistical significance. Computerized physician order entry and clinical practice guideline systems comprise the greatest number of evaluated CDS tools (13 and 10 respectively), followed by those that utilize patient-reported outcomes (8), clinical pathway systems (8) and prescriber alerts for best-practice advisories (4). Our review indicates that CDS can improve guideline adherence, patient-centered care, and care delivery processes in oncology.


Subject(s)
Decision Support Systems, Clinical , Medical Order Entry Systems , Humans , Medical Oncology
5.
J Appl Clin Med Phys ; 24(5): e13916, 2023 May.
Article in English | MEDLINE | ID: mdl-36763085

ABSTRACT

PURPOSE: Adaptive radiation therapy (ART) on the integrated Elekta Unity magnetic resonance (MR)-linac requires routine quality assurance to verify delivery accuracy and system data transfer. In this work, our objective was to develop and validate a novel automated end-to-end test suite that verifies data transfer between multiple software platforms and quantifies the performance of multiple machine subcomponents critical to the ART process. METHODS: We designed and implemented a software tool to quantify the MR and megavoltage (MV) isocenter coincidence, treatment couch positioning consistency, isocenter shift accuracy for the adapted plan as well as the MLC and jaw position accuracy following the beam aperture adaptation. Our tool employs a reference treatment plan with a simulated isocenter shift generated on an MR image of a readily available phantom with MR and MV visible fiducials. Execution of the test occurs within the standard adapt-to-position (ATP) clinical workflow with MV images collected of the delivered treatment fields. Using descriptive statistics, we quantified uncertainty in couch positioning, isocentre shift as well as the jaw and MLC positions of the adapted fields. We also executed sensitivity measurements to evaluate the detection algorithm's performance. RESULTS: We report the results of 301 daily testing instances. We demonstrated consistent tracking of the MR-to-MV alignment with respect to the established value and to detect small changes on the order of 0.2 mm following machine service events. We found couch position consistency relative to the test baseline value was within 95% CI [-0.31, 0.26 mm]. For phantom shifts that form the basis for the plan adaptation, we found agreement between MV-image-detected phantom shift and online image registration, within ± 1.5 mm in all directions with a 95% CI difference of [-1.29, 0.79 mm]. For beam aperture adaptation accuracy, we found differences between the planned and detected jaw positions had a mean value of 0.27 mm and 95% CI of [-0.29, 0.82 mm] and -0.17 mm and 95% CI of [-0.37, 0.05 mm] for the MLC positions. Automated fiducial detected accuracy was within 0.08 ± 0.20 mm of manual localization. Introduced jaw and MLC position errors (1-10 mm) were detected within 0.55 mm (within 1 mm for 15/256 instances for the jaws). Phantom shifts (1.3 or 5 mm in each cardinal direction) from a reference position were detected within 0.26 mm. CONCLUSIONS: We have demonstrated the accuracy and sensitivity of a daily end-to-end test suite capable of detecting errors in multiple machine subcomponents including system data transfer. Our test suite evaluates the entire treatment workflow and has captured system communication issues prior to patient treatment. With automated processing and the use of a standard vendor-provided phantom, it is possible to expand to other Unity sites.


Subject(s)
Magnetic Resonance Imaging , Particle Accelerators , Humans , Magnetic Resonance Imaging/methods , Software , Quality Control , Phantoms, Imaging , Magnetic Resonance Spectroscopy , Radiotherapy Planning, Computer-Assisted/methods
6.
Technol Cancer Res Treat ; 21: 15330338221123108, 2022.
Article in English | MEDLINE | ID: mdl-36285543

ABSTRACT

ObjectivesPandemics, natural disasters, and other unforeseen circumstances can cause short-term variation in radiotherapy utilization. In this study, we aim to develop a model to forecast linear accelerator (LINAC) utilization during periods of varying workloads. Methods: Using computed tomography (CT)-simulation data and the rate of new LINAC appointment bookings in the preceding week as input parameters, a multiple linear regression model to forecast LINAC utilization over a 15-working day horizon was developed and tested on institutional data. Results: Future LINAC utilization was estimated in our training dataset with a forecasting error of 3.3%, 5.9%, and 7.2% on days 5, 10, and 15, respectively. The model identified significant variations (≥5% absolute differences) in LINAC utilization with an accuracy of 69%, 62%, and 60% on days 5, 10, and 15, respectively. The results were similar in the validation dataset with forecasting errors of 3.4%, 5.3%, and 6.2% and accuracy of 67%, 60%, and 58% on days 5, 10, and 15, respectively. These results compared favorably to moving average and exponential smoothing forecasting techniques. Conclusions: The developed linear regression model was able to accurately forecast future LINAC utilization based on LINAC booking rate and CT simulation data, and has been incorporated into our institutional dashboard for broad distribution. Advances in knowledge: Our proposed linear regression model is a practical and intuitive approach to forecasting short-term LINAC utilization, which can be used for resource planning and allocation during periods with varying LINAC workloads.


Subject(s)
Particle Accelerators , Workload , Humans , Forecasting , Linear Models
7.
Phys Med Biol ; 67(11)2022 06 02.
Article in English | MEDLINE | ID: mdl-35576920

ABSTRACT

Purpose.It has been postulated that the delivery of radiotherapy at ultra-high dose rates ('FLASH') reduces normal tissue toxicities by depleting them of oxygen. The fraction of normal tissue and cancer cells surviving radiotherapy depends on dose and oxygen levels in an exponential manner and even a very small fraction of tissue at low oxygen levels can determine radiotherapy response. To quantify the differential impact of FLASH radiotherapy on normal and tumour tissues, the spatial heterogeneity of oxygenation in tissue should thus be accounted for.Methods.The effect of FLASH on radiation-induced normal and tumour tissue cell killing was studied by simulating oxygen diffusion, metabolism, and radiolytic oxygen depletion (ROD) over domains with simulated capillary architectures. To study the impact of heterogeneity, two architectural models were used: (1) randomly distributed capillaries and (2) capillaries forming a regular square lattice array. The resulting oxygen partial pressure distribution histograms were used to simulate normal and tumour tissue cell survival using the linear quadratic model of cell survival, modified to incorporate oxygen-enhancement ratio effects. The ratio ('dose modifying factors') of conventional low-dose-rate dose and FLASH dose at iso-cell survival was computed and compared with empirical iso-toxicity dose ratios.Results.Tumour cell survival was found to be increased by FLASH as compared to conventional radiotherapy, with a 0-1 order of magnitude increase for expected levels of tumour hypoxia, depending on the relative magnitudes of ROD and tissue oxygen metabolism. Interestingly, for the random capillary model, the impact of FLASH on well-oxygenated (normal) tissues was found to be much greater, with an estimated increase in cell survival by up to 10 orders of magnitude, even though reductions in mean tissue partial pressure were modest, less than ∼7 mmHg for the parameter values studied. The dose modifying factor for normal tissues was found to lie in the range 1.2-1.7 for a representative value of normal tissue oxygen metabolic rate, consistent with preclinical iso-toxicity results.Conclusions.The presence of very small nearly hypoxic regions in otherwise well-perfused normal tissues with high mean oxygen levels resulted in a greater proportional sparing of normal tissue than tumour cells during FLASH irradiation, possibly explaining empirical normal tissue sparing and iso-tumour control results.


Subject(s)
Neoplasms , Radiation Oncology , Cell Survival , Humans , Neoplasms/radiotherapy , Oxygen/metabolism , Radiotherapy , Radiotherapy Dosage
8.
Front Oncol ; 12: 1086258, 2022.
Article in English | MEDLINE | ID: mdl-36776378

ABSTRACT

MRI-linear accelerator (MR-linac) devices have been introduced into clinical practice in recent years and have enabled MR-guided adaptive radiation therapy (MRgART). However, by accounting for anatomical changes throughout radiation therapy (RT) and delivering different treatment plans at each fraction, adaptive radiation therapy (ART) highlights several challenges in terms of calculating the total delivered dose. Dose accumulation strategies-which typically involve deformable image registration between planning images, deformable dose mapping, and voxel-wise dose summation-can be employed for ART to estimate the delivered dose. In MRgART, plan adaptation on MRI instead of CT necessitates additional considerations in the dose accumulation process because MRI pixel values do not contain the quantitative information used for dose calculation. In this review, we discuss considerations for dose accumulation specific to MRgART and in relation to current MR-linac clinical workflows. We present a general dose accumulation framework for MRgART and discuss relevant quality assurance criteria. Finally, we highlight the clinical importance of dose accumulation in the ART era as well as the possible ways in which dose accumulation can transform clinical practice and improve our ability to deliver personalized RT.

9.
Phys Med Biol ; 66(18)2021 09 09.
Article in English | MEDLINE | ID: mdl-34438383

ABSTRACT

Dose painting of hypoxic tumour sub-volumes using positron-emission tomography (PET) has been shown to improve tumour controlin silicoin several sites, predominantly head and neck and lung cancers. Pancreatic cancer presents a more stringent challenge, given its proximity to critical gastro-intestinal organs-at-risk (OARs), anatomic motion, and impediments to reliable PET hypoxia quantification. A radiobiological model was developed to estimate clonogen survival fraction (SF), using18F-fluoroazomycin arabinoside PET (FAZA PET) images from ten patients with unresectable pancreatic ductal adenocarcinoma to quantify oxygen enhancement effects. For each patient, four simulated five-fraction stereotactic body radiotherapy (SBRT) plans were generated: (1) a standard SBRT plan aiming to cover the planning target volume with 40 Gy, (2) dose painting plans delivering escalated doses to a maximum of three FAZA-avid hypoxic sub-volumes, (3) dose painting plans with simulated spacer separating the duodenum and pancreatic head, and (4), plans with integrated boosts to geometric contractions of the gross tumour volume (GTV). All plans saturated at least one OAR dose limit. SF was calculated for each plan and sensitivity of SF to simulated hypoxia quantification errors was evaluated. Dose painting resulted in a 55% reduction in SF as compared to standard SBRT; 78% with spacer. Integrated boosts to hypoxia-blind geometric contractions resulted in a 41% reduction in SF. The reduction in SF for dose-painting plans persisted for all hypoxia quantification parameters studied, including registration and rigid motion errors that resulted in shifts and rotations of the GTV and hypoxic sub-volumes by as much as 1 cm and 10 degrees. Although proximity to OARs ultimately limited dose escalation, with estimated SFs (∼10-5) well above levels required to completely ablate a ∼10 cm3tumour, dose painting robustly reduced clonogen survival when accounting for expected treatment and imaging uncertainties and thus, may improve local response and associated morbidity.


Subject(s)
Pancreatic Neoplasms , Radiosurgery , Radiotherapy, Intensity-Modulated , Humans , Hypoxia , Pancreatic Neoplasms/diagnostic imaging , Pancreatic Neoplasms/radiotherapy , Positron-Emission Tomography , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Tomography, X-Ray Computed
10.
Med Phys ; 48(10): e830-e885, 2021 Oct.
Article in English | MEDLINE | ID: mdl-34036590

ABSTRACT

The charges on this task group (TG) were as follows: (a) provide specific procedural guidelines for performing the tests recommended in TG 142; (b) provide estimate of the range of time, appropriate personnel, and qualifications necessary to complete the tests in TG 142; and (c) provide sample daily, weekly, monthly, or annual quality assurance (QA) forms. Many of the guidelines in this report are drawn from the literature and are included in the references. When literature was not available, specific test methods reflect the experiences of the TG members (e.g., a test method for door interlock is self-evident with no literature necessary). In other cases, the technology is so new that no literature for test methods was available. Given broad clinical adaptation of volumetric modulated arc therapy (VMAT), which is not a specific topic of TG 142, several tests and criteria specific to VMAT were added.


Subject(s)
Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Particle Accelerators , Quality Assurance, Health Care , Radiotherapy Dosage
11.
Cureus ; 13(2): e13606, 2021 Feb 28.
Article in English | MEDLINE | ID: mdl-33816005

ABSTRACT

Stereotactic radiotherapy (SBRT) has been applied to treat cardiac arrhythmias, but our institution had not yet implemented this technique. Here, we explain how we used implementation science and knowledge translation to provide cardiac SBRT to a critically ill patient with malignancy-associated refractory ventricular tachycardia. We reviewed the critical factors that enabled the implementation of this urgent treatment, such as the context of the implementation, the characteristics of the intervention, and the stakeholders. These principles can be used by other radiation programs to implement novel treatments in urgent settings, where the gold standard process of planning and developing policies and protocols is not possible.

12.
Acta Oncol ; 60(2): 260-266, 2021 Feb.
Article in English | MEDLINE | ID: mdl-33170058

ABSTRACT

INTRODUCTION: Liver cancers are challenging to treat using image-guided radiotherapy (IGRT) due to motion and deformation of target volumes and organs at risk (OARs), as well as difficulties in visualising liver tumours using cone-beam computed tomography (CBCT) based IGRT. Liver cancer patients may thus benefit from magnetic resonance (MR)-guided daily adaptive re-planning. We evaluated the dosimetric impact of a daily plan adaptation strategy based on daily MR imaging versus CBCT-based IGRT. METHODS: Ten patients were studied who were treated with CBCT-guided five-fraction stereotactic body radiotherapy (SBRT) and underwent MR imaging before each fraction. Simulated reference plans were created on computer tomography (CT) images and adapted plans were created on the daily MR images. Two plan adaptation strategies were retrospectively simulated: (1) translational couch shifts to match liver, mimicking standard CBCT guidance and (2) daily plan adaptation based on reference plan clinical goals and daily target and OAR contours. Dose statistics were calculated for both strategies and compared. RESULTS: Couch shifts resulted in an average reduction in GTV D99% relative to reference plan values of 5.2 Gy (-12.5% of reference values). Daily plan adaptation reduced this to 0.8 Gy (-2.0%). For six patients who were OAR dose-limited on reference plans, couch shifts resulted in OAR dose violations in 28 out of 28 simulated fractions, respectively; no violations occurred using daily plan adaptation. No OAR dose violations occurred using either strategy for the four cases not OAR dose-limited at reference planning. CONCLUSIONS: MR-guided daily plan adaptation ensured OAR dose constraints were met at all simulated treatment fractions while CBCT-based IGRT resulted in a systematic over-dosing of OARs in patients whose doses were limited by OAR dose at the time of reference planning.


Subject(s)
Radiosurgery , Radiotherapy, Image-Guided , Radiotherapy, Intensity-Modulated , Humans , Liver/diagnostic imaging , Magnetic Resonance Spectroscopy , Organs at Risk , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Retrospective Studies
13.
J Appl Clin Med Phys ; 20(7): 100-108, 2019 Jul.
Article in English | MEDLINE | ID: mdl-31199568

ABSTRACT

PURPOSE: To evaluate the performance and stability of Elekta Agility multi-leaf collimator (MLC) leaf positioning using a daily, automated quality control (QC) test based on megavoltage (MV) images in combination with statistical process control tools, and identify special causes of variations in performance. METHODS: Leaf positions were collected daily for 13 Elekta linear accelerators over 11-37 months using the automated QC test, which analyzes 23 MV images to determine the location of MLC leaves relative to radiation isocenter. Leaf positioning stability was assessed using individual and moving range control charts. Specification levels of ±0.5, ±1, and ±1.5 mm were tested to determine positional accuracy. The durations between out-of-control and out-of-specification events were determined. Peaks in out-of-control leaf positions were identified and correlated to servicing events recorded for the whole duration of data collection. RESULTS: Mean leaf position error was -0.01 mm (range -1.3-1.6). Data stayed within ±1 mm specification for 457 days on average (range 3-838) and within ±1.5 mm for the entire date range. Measurements stayed within ±0.5 mm for 1 day on average (range 0-17); however, our MLC leaves were not calibrated to this level of accuracy. Leaf position varied little over time, as confirmed by tight individual (mean ±0.19 mm, range 0.09-0.43) and moving range (mean 0.23 mm, range 0.10-0.53) control limits. Due to sporadic out-of-control events, the mean in-control duration was 2.8 days (range 1-28.5). A number of factors were found to contribute to leaf position errors and out-of-control behavior, including servicing events, beam spot motion, and image artifacts. CONCLUSIONS: The Elekta Agility MLC model was found to perform with high stability, as evidenced by the tight control limits. The in-specification durations support the current recommendation of monthly MLC QC tests with a ±1 mm tolerance. Future work is on-going to determine if performance can be optimized further using high-frequency QC test results to drive recalibration frequency.


Subject(s)
Models, Statistical , Particle Accelerators/instrumentation , Quality Control , Radiotherapy Planning, Computer-Assisted/methods , Calibration , Humans , Radiotherapy Dosage , Radiotherapy, Intensity-Modulated/methods
14.
J Med Imaging Radiat Sci ; 50(2): 297-307, 2019 06.
Article in English | MEDLINE | ID: mdl-31176438

ABSTRACT

BACKGROUND: CT simulator for radiation therapy aims to produce high-quality images for dose calculation and delineation of target and organs at risk in the process of treatment planning. Selection of CT imaging protocols that achieve a desired image quality while minimizing patient dose depends on technical CT parameters and their relationship with image quality and radiation dose. For similar imaging protocols using comparable technical CT parameters, there are also variations in image quality metrics between different CT simulator models. Understanding the relationship and variation is important for selecting appropriate imaging protocol and standardizing QC process. Here, we proposed an automated method to determine the relationship between image quality and radiation dose for various CT technical parameters. MATERIAL AND METHOD: The impact of scan parameters on various aspects of image quality and volumetric CT dose index for a Philips Brilliance Big Bore and a Toshiba Aquilion One CT scanners were determined by using commercial phantom and automated image quality analysis software and cylindrical radiation dose phantom. RESULTS AND DISCUSSION: Both scanners had very similar and satisfactory performance based on the diagnostic acceptance criteria recommended by ACR, International Atomic Energy Agency, and American Association of Physicists in Medicine. However, our results showed a compromise between different image quality components such as low-contrast and spatial resolution with the change of scanning parameters and revealed variations between the two scanners on their image quality performance. Measurement using a generic phantom and analysis by automated software was unbiased and efficient. CONCLUSION: This method provides information that can be used as a baseline for CT scanner image quality and dosimetric QC for different CT scanner models in a given institution or across sites.


Subject(s)
Radiation Dosage , Radiotherapy Planning, Computer-Assisted , Tomography Scanners, X-Ray Computed/standards , Tomography, X-Ray Computed/instrumentation , Algorithms , Computer Simulation , Humans , Phantoms, Imaging , Quality Control , Radiotherapy Planning, Computer-Assisted/instrumentation , Radiotherapy Planning, Computer-Assisted/standards
15.
J Appl Clin Med Phys ; 20(3): 81-88, 2019 Mar.
Article in English | MEDLINE | ID: mdl-30817079

ABSTRACT

PURPOSE: Seasonal trends in linear accelerator output have been reported by at least one institution and data have suggested that they may be present at our center as well. The purpose of this work was to characterize these trends and determine whether local environmental conditions within the treatment rooms may be impacting the linear accelerators and/or the quality control (QC) dosimeter. METHODS: Runtime plots of daily output data, acquired using an in-house ion chamber-based device, over 3 yr and for 15 linear accelerators of different makes and models were reviewed and evaluated. Environmental conditions were monitored prospectively in a representative treatment room for approximately 9 months and evaluated for correlations with output trends. Independent measures of output using daily MV portal images were compared with output measurements using the ion chamber-based device. A separate controlled experiment probing the response of the in-house dosimeter to humidity changes over time was also carried out using a constant current source and a small enclosure. RESULTS: Runtime plots of output revealed sinusoidal, seasonal variations that were consistent across all treatment units, irrespective of manufacturer, model, or age of machine. The amplitude of the variation was on the order of 1% and maintained a yearly period. The independent measure of output using MV portal images did not corroborate the seasonal trends observed with the daily QC dosimeter. Based on the controlled experiment, the QC dosimeter was found to have a dependence on relative humidity changes, decreasing 1% in output per 30% increase in relative humidity. CONCLUSIONS: Results confirm the presence of underlying seasonal variations in measured output from the linear accelerators. The findings identify humidity impact on the measurement device as the underlying cause of the cyclical changes and not the accelerators themselves. These results could help minimize unwarranted machine servicing.


Subject(s)
Calibration , Particle Accelerators/instrumentation , Particle Accelerators/standards , Radiometry/instrumentation , Seasons , Humidity , Photons
16.
Neurosurgery ; 84(2): 435-441, 2019 02 01.
Article in English | MEDLINE | ID: mdl-29547929

ABSTRACT

BACKGROUND: Stereotactic body radiotherapy (SBRT) of the spine provides superior tumor control, but vertebral compression fractures are increased and the pathophysiological process underneath is not well understood. Data on histopathological changes, particularly after salvage SBRT (sSBRT) following conventional irradiation, are scarce. OBJECTIVE: To investigate surgical specimens after sSBRT and primary SBRT (pSBRT) regarding histopathological changes. METHODS: We assessed 704 patients treated with spine SBRT 2006 to 2012. Thirty patients underwent salvage surgery; 23 histopathological reports were available. Clinical and histopathological findings were analyzed for sSBRT (69.6%) and pSBRT (30.4%). RESULTS: Mean time to surgery after sSBRT/pSBRT was 8.3/10.3 mo (P = .64). Reason for surgery included pain (sSBRT/pSBRT: 12.5%/71.4%, P = .25), fractures (sSBRT/pSBRT: 37.5%/28.6%, P = .68), and neurological symptoms (sSBRT/pSBRT: 68.8%/42.9%, P = .24). Radiological tumor progression after sSBRT/pSBRT was seen in 71.4%/42.9% (P = .2). Most specimens displayed viable/proliferative tumor (sSBRT/pSBRT: 62.5%/71.4%, P = .68 and 56.3%/57.1%, P = .97). Few specimens showed soft tissue necrosis (sSBRT/pSBRT: 20%/28.6%, P = .66), osteonecrosis (sSBRT/pSBRT: 14.3%/16.7%, P = .89), or bone marrow fibrosis (sSBRT/pSBRT: 42.9%/33.3%, P = .69). Tumor bed necrosis was more common after sSBRT (81.3%/42.9%, P = .066). Radiological tumor progression correlated with viable/proliferative tumor (P = .03/P = .006) and tumor bed necrosis (P = .03). Fractures were increased with bone marrow fibrosis (P = .07), but not with osteonecrosis (P = .53) or soft tissue necrosis (P = .19). Neurological symptoms were common with radiological tumor progression (P = .07), but not with fractures (P = .18). CONCLUSION: For both, sSBRT and pSBRT, histopathological changes were similar. Neurological symptoms were attributable to tumor progression and pathological fractures were not associated with osteonecrosis or tumor progression.


Subject(s)
Radiosurgery/adverse effects , Re-Irradiation/adverse effects , Salvage Therapy/adverse effects , Spinal Neoplasms/radiotherapy , Adult , Aged , Cohort Studies , Female , Fractures, Compression/epidemiology , Fractures, Compression/etiology , Humans , Male , Middle Aged , Necrosis/epidemiology , Necrosis/etiology , Radiation Injuries/epidemiology , Radiation Injuries/etiology , Radiosurgery/methods , Re-Irradiation/methods , Salvage Therapy/methods , Spinal Fractures/epidemiology , Spinal Fractures/etiology , Spinal Neoplasms/secondary
17.
Med Phys ; 45(6): 2388-2398, 2018 Jun.
Article in English | MEDLINE | ID: mdl-29645282

ABSTRACT

PURPOSE: The delivery accuracy of highly conformal dose distributions generated using intensity modulation and collimator, gantry, and couch degrees of freedom is directly affected by the quality of the alignment between the radiation beam and the mechanical axes of a linear accelerator. For this purpose, quality control (QC) guidelines recommend a tolerance of ±1 mm for the coincidence of the radiation and mechanical isocenters. Traditional QC methods for assessment of radiation and mechanical axes alignment (based on pointer alignment) are time consuming and complex tasks that provide limited accuracy. In this work, an automated test suite based on an analytical model of the linear accelerator motions was developed to streamline the QC of radiation and mechanical axes alignment. METHODS: The proposed method used the automated analysis of megavoltage images of two simple task-specific phantoms acquired at different linear accelerator settings to determine the coincidence of the radiation and mechanical isocenters. The sensitivity and accuracy of the test suite were validated by introducing actual misalignments on a linear accelerator between the radiation axis and the mechanical axes using both beam steering and mechanical adjustments of the gantry and couch. RESULTS: The validation demonstrated that the new QC method can detect sub-millimeter misalignment between the radiation axis and the three mechanical axes of rotation. A displacement of the radiation source of 0.2 mm using beam steering parameters was easily detectable with the proposed collimator rotation axis test. Mechanical misalignments of the gantry and couch rotation axes of the same magnitude (0.2 mm) were also detectable using the new gantry and couch rotation axis tests. For the couch rotation axis, the phantom and test design allow detection of both translational and tilt misalignments with the radiation beam axis. For the collimator rotation axis, the test can isolate the misalignment between the beam radiation axis and the mechanical collimator rotation axis from the impact of field size asymmetry. The test suite can be performed in a reasonable time (30-35 min) due to simple phantom setup, prescription-based beam delivery, and automated image analysis. As well, it provides a clear description of the relationship between axes. After testing the sensitivity of the test suite to beam steering and mechanical errors, the results of the test suite were used to reduce the misalignment errors of the linac to less than 0.7-mm radius for all axes. CONCLUSIONS: The proposed test suite offers sub-millimeter assessment of the coincidence of the radiation and mechanical isocenters and the test automation reduces complexity with improved efficiency. The test suite results can be used to optimize the linear accelerator's radiation to mechanical isocenter alignment by beam steering and mechanical adjustment of gantry and couch.


Subject(s)
Particle Accelerators , Quality Control , Radiotherapy, Conformal/instrumentation , Radiotherapy, Conformal/methods , Automation , Equipment Design , Models, Theoretical , Phantoms, Imaging , Rotation
18.
Radiat Oncol ; 12(1): 153, 2017 Sep 11.
Article in English | MEDLINE | ID: mdl-28893299

ABSTRACT

OBJECT: Stereotactic body radiotherapy (SBRT) for vertebral metastases has emerged as a promising technique, offering high rates of symptom relief and local control combined with low risk of toxicity. Nonetheless, local failure or vertebral instability may occur after spine SBRT, generating the need for subsequent surgery in the irradiated region. This study evaluated whether there is an increased incidence of surgical complications in patients previously treated with SBRT at the index level. METHODS: Based upon a retrospective international database of 704 cases treated with SBRT for vertebral metastases, 30 patients treated at 6 different institutions were identified who underwent surgery in a region previously treated with SBRT. RESULTS: Thirty patients, median age 59 years (range 27-84 years) underwent SBRT for 32 vertebral metastases followed by surgery at the same vertebra. Median follow-up time from SBRT was 17 months. In 17 cases, conventional radiotherapy had been delivered prior to SBRT at a median dose of 30 Gy in median 10 fractions. SBRT was administered with a median prescription dose of 19.3 Gy (range 15-65 Gy) delivered in median 1 fraction (range 1-17) (median EQD2/10 = 44 Gy). The median time interval between SBRT and surgical salvage therapy was 6 months (range 1-39 months). Reasons for subsequent surgery were pain (n = 28), neurological deterioration (n = 15) or fracture of the vertebral body (n = 13). Open surgical decompression (n = 24) and/or stabilization (n = 18) were most frequently performed; Five patients (6 vertebrae) were treated without complications with vertebroplasty only. Increased fibrosis complicating the surgical procedure was explicitly stated in one surgical report. Two durotomies occurred which were closed during the operation, associated with a neurological deficit in one patient. Median blood loss was 500 ml, but five patients had a blood loss of more than 1 l during the procedure. Delayed wound healing was reported in two cases. One patient died within 30 days of the operation. CONCLUSION: In this series of surgical interventions following spine SBRT, the overall complication rate was 19%, which appears comparable to primary surgery without previous SBRT. Prior spine SBRT does not appear to significantly increase the risk of intra- and post-surgical complications.


Subject(s)
Postoperative Complications/epidemiology , Radiosurgery/adverse effects , Spinal Neoplasms/radiotherapy , Spine/radiation effects , Spine/surgery , Adult , Aged , Aged, 80 and over , Female , Humans , Incidence , Male , Middle Aged , Orthopedic Procedures/adverse effects , Retrospective Studies , Risk Factors , Spinal Neoplasms/secondary
19.
J Neurosurg Spine ; 25(5): 646-653, 2016 Nov.
Article in English | MEDLINE | ID: mdl-27341054

ABSTRACT

OBJECTIVE This study is a multi-institutional pooled analysis specific to imaging-based local control of spinal metastases in patients previously treated with conventional external beam radiation therapy (cEBRT) and then treated with re-irradiation stereotactic body radiotherapy (SBRT) to the spine as salvage therapy, the largest such study to date. METHODS The authors reviewed cases involving 215 patients with 247 spinal target volumes treated at 7 institutions. Overall survival was calculated on a patient basis, while local control was calculated based on the spinal target volume treated, both using the Kaplan-Meier method. Local control was defined as imaging-based progression within the SBRT target volume. Equivalent dose in 2-Gy fractions (EQD2) was calculated for the cEBRT and SBRT course using an α/ß of 10 for tumor and 2 for both spinal cord and cauda equina. RESULTS The median total dose/number of fractions of the initial cEBRT was 30 Gy/10. The median SBRT total dose and number of fractions were 18 Gy and 1, respectively. Sixty percent of spinal target volumes were treated with single-fraction SBRT (median, 16.6 Gy and EQD2/10 = 36.8 Gy), and 40% with multiple-fraction SBRT (median 24 Gy in 3 fractions, EQD2/10 = 36 Gy). The median time interval from cEBRT to re-irradiation SBRT was 13.5 months, and the median duration of patient follow-up was 8.1 months. Kaplan-Meier estimates of 6- and 12-month overall survival rates were 64% and 48%, respectively; 13% of patients suffered a local failure, and the 6- and 12-month local control rates were 93% and 83%, respectively. Multivariate analysis identified Karnofsky Performance Status (KPS) < 70 as a significant prognostic factor for worse overall survival, and single-fraction SBRT as a significant predictive factor for better local control. There were no cases of radiation myelopathy, and the vertebral compression fracture rate was 4.5%. CONCLUSIONS Re-irradiation spine SBRT is effective in yielding imaging-based local control with a clinically acceptable safety profile. A randomized trial would be required to determine the optimal fractionation.


Subject(s)
Radiosurgery/methods , Re-Irradiation/methods , Salvage Therapy/methods , Spinal Neoplasms/secondary , Spinal Neoplasms/surgery , Adolescent , Adult , Aged , Aged, 80 and over , Cohort Studies , Female , Follow-Up Studies , Humans , Kaplan-Meier Estimate , Male , Middle Aged , Radiosurgery/adverse effects , Radiotherapy Dosage , Re-Irradiation/adverse effects , Salvage Therapy/adverse effects , Spinal Neoplasms/mortality , Spine/surgery , Time Factors , Treatment Outcome , Young Adult
20.
Radiat Oncol ; 11: 57, 2016 Apr 18.
Article in English | MEDLINE | ID: mdl-27089966

ABSTRACT

BACKGROUND: The aim of this study was to quantify the variability in spinal radiosurgery (SRS) planning practices between five international institutions, all member of the Elekta Spine Radiosurgery Research Consortium. METHODS: Four institutions provided one representative patient case each consisting of the medical history, CT and MR imaging. A step-wise planning approach was used where, after each planning step a consensus was generated that formed the basis for the next planning step. This allowed independent analysis of all planning steps of CT-MR image registration, GTV definition, CTV definition, PTV definition and SRS treatment planning. In addition, each institution generated one additional SRS plan for each case based on intra-institutional image registration and contouring, independent of consensus results. RESULTS: Averaged over the four cases, image registration variability ranged between translational 1.1 mm and 2.4 mm and rotational 1.1° and 2.0° in all three directions. GTV delineation variability was 1.5 mm in axial and 1.6 mm in longitudinal direction averaged for the four cases. CTV delineation variability was 0.8 mm in axial and 1.2 mm in longitudinal direction. CTV-to-PTV margins ranged between 0 mm and 2 mm according to institutional protocol. Delineation variability was 1 mm in axial directions for the spinal cord. Average PTV coverage for a single fraction18 Gy prescription was 87 ± 5 %; Dmin to the PTV was 7.5 ± 1.8 Gy averaged over all cases and institutions. Average Dmax to the PRV_SC (spinal cord + 1 mm) was 10.5 ± 1.6 Gy and the average Paddick conformity index was 0.69 ± 0.06. CONCLUSIONS: Results of this study reflect the variability in current practice of spine radiosurgery in large and highly experienced academic centers. Despite close methodical agreement in the daily workflow, clinically significant variability in all steps of the treatment planning process was demonstrated. This may translate into differences in patient clinical outcome and highlights the need for consensus and established delineation and planning criteria.


Subject(s)
Radiosurgery/methods , Spinal Neoplasms/radiotherapy , Female , Humans , Image Processing, Computer-Assisted , International Cooperation , Magnetic Resonance Imaging , Male , Middle Aged , Multimodal Imaging , Organs at Risk , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Reproducibility of Results , Spinal Cord/radiation effects , Tomography, X-Ray Computed , Treatment Outcome
SELECTION OF CITATIONS
SEARCH DETAIL
...