J Med Assoc Thai ; 98(1): 82-7, 2015 Jan.
Article in English | MEDLINE | ID: mdl-25775737

ABSTRACT

BACKGROUND: Hand external fixators are in use worldwide and insulin syringes can generally be found in an operating room. OBJECTIVE: To compare the pull-out strength between degrees of Kirschner wire fixation (0° and 30°) and the effect of filling an insulin syringe with polymer MATERIAL AND METHOD: Pull-out strength was compared between a syringe externalfixator and a bone or plastic tube model. Fifty-two plastic tube models and 20 dry phalangeal bones were included The syringe external fixator was attached via two Kirschner wires. Four variations were studied: 0° non-polymer 0° with polymer augmentation, 30° non-polymer and 30° with polymer augmentation. The pull-out strength was measured in each group. RESULTS: The strength of polymer augmentation was higher than non-polymer augmentation at 0° (p = 0.0003) and 30° (p = 0.0002). The Kirschner wire at 30° provided more pull-out strength than at 0° (p = 0.0003) using the syringe with no polymer However, using the syringe with polymer augmentation, there was no significant difference (p = 0.5136). CONCLUSION: Polymer augmentation significantly increases the pull-out strength at both degrees offixation. The degree of fixation significantly increases the pull-out strength only in the non-polymer group, where pinning at 30° was superior to 0°.


Subject(s)
Bone Wires , External Fixators , Fracture Fixation/methods , Syringes , Tensile Strength , Biomechanical Phenomena , Hand Injuries/surgery , Humans , Materials Testing , Models, Biological , Polymers
SELECTION OF CITATIONS
SEARCH DETAIL
...