Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Int J Med Robot ; 16(1): e2050, 2020 Feb.
Article in English | MEDLINE | ID: mdl-31677219

ABSTRACT

BACKGROUND: Manipulation of the endoscope during minimally invasive surgery is a major source of inconvenience and discomfort. This report elucidates the architecture of a novel one-hand controlled endoscope positioning device and presents a practicability evaluation. METHODS AND MATERIALS: Setup time and total surgery time, number and duration of the manipulations, side effects of three-dimensional (3D) imaging, and ergonomic complaints were assessed by three surgeons during cadaveric and in vivo porcine trials. RESULTS: Setup was accomplished in an average (SD) of 230 (120) seconds. The manipulation time was 3.87 (1.77) seconds for angular movements and 0.83 (0.24) seconds for zooming, with an average (SD) of 30.5 (16.3) manipulations per procedure. No side effects of 3D imaging or ergonomic complaints were reported. CONCLUSIONS: The integration of an active zoom into a passive endoscope holder delivers a convenient synergy between a human and a machine-controlled holding device. It is shown to be safe, simple, and intuitive to use and allows unrestrained autonomic control of the endoscope by the surgeon.


Subject(s)
Endoscopes , Equipment Design , Animals , Feasibility Studies , Humans , Minimally Invasive Surgical Procedures/instrumentation , Surgeons , Swine
SELECTION OF CITATIONS
SEARCH DETAIL
...