1.
Rev Sci Instrum ; 89(1): 015110, 2018 Jan.
Article in English | MEDLINE | ID: mdl-29390699

ABSTRACT

The characterization of nanostructured samples with at least one restricted dimension like thin films or nanowires is challenging, but important to understand their structure and transport mechanism, and to improve current industrial products and production processes. We report on the 2nd generation of a measurement chip, which allows for a simplified sample preparation process, and the measurement of samples deposited from the liquid phase using techniques like spin coating and drop casting. The new design enables us to apply much higher temperature gradients for the Seebeck coefficient measurement in a shorter time, without influencing the sample holder's temperature distribution. Furthermore, a two membrane correction method for the 3ω thermal conductivity measurement will be presented, which takes the heat loss due to radiation into account and increases the accuracy of the measurement results significantly. Errors caused by different sample compositions, varying sample geometries, and different heat profiles are avoided with the presented measurement method. As a showcase study displaying the validity and accuracy of our platform, we present temperature-dependent measurements of the thermoelectric properties of an 84 nm Bi87Sb13 thin film and a 15 µm PEDOT:PSS thin film.