ABSTRACT
PURPOSE: To evaluate retrievability of the Celect vena cava filter over time and to assess the safety of the retrieval procedure in a prospective multicenter registry. MATERIALS AND METHODS: Between October 2005 and March 2008, Celect filters were placed in 95 patients (61 men; mean age, 51 years +/- 18.5) with a temporary need for an inferior vena cava (IVC) filter. All patients satisfied requirements for filter placement; the primary indications for placement were pulmonary embolism (PE) with a contraindication to or failure of anticoagulation (n = 40), high risk for further PE (n = 29), trauma (n = 23), or massive PE with residual deep vein thrombosis and risk for further PE (n = 3). Filter orientation, vena cava injury, and other device-related incidents were evaluated at implantation and retrieval. The degree of difficulty associated with retrieval was also assessed. RESULTS: Filter retrieval was attempted in 58 patients (mean indwell time of 179 days; median, 168.5 d; range, 5-466 d). Fifty-six filters (96.6%) were successfully retrieved. Unsuccessful retrieval attempts were attributed to filter tilt (n = 1) or excessive tissue growth with the hook embedded in the endothelium (n = 1). No adverse events were associated with the inability to retrieve these filters. A Kaplan-Meier product-limit survival estimate revealed that the probability of successful filter retrieval remained at 100% at 50 weeks and at more than 74% at 55 weeks after implantation. No adverse events were related to the retrieval procedures. CONCLUSIONS: Retrieval of the Celect filter was performed safely as long as 466 days after implantation.