Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
JBJS Case Connect ; 11(2)2021 06 24.
Article in English | MEDLINE | ID: mdl-34166254

ABSTRACT

CASE: We describe 2 cases of gross trunnion failure (GTF) in males with Centers for Disease Control and Prevention Class 3 obesity 10 years after metal-on-polyethylene total hip arthroplasty (THA) with a titanium-alloy femoral stem and Type 1 taper. One patient received a large diameter cobalt-chromium femoral head, whereas the other received a smaller diameter head, both with high-offset femoral stems. CONCLUSION: This is the first report of GTF involving the Echo Bi-Metric femoral stem after metal-on-polyethylene THA, and surgeons should consider the potential complication of GTF when using this specific femoral stem with metal heads in obese male patients.


Subject(s)
Arthroplasty, Replacement, Hip , Hip Prosthesis , Arthroplasty, Replacement, Hip/adverse effects , Hip Prosthesis/adverse effects , Humans , Male , Polyethylene , Prosthesis Design , Prosthesis Failure , United States