J Crit Care ; 67: 21-25, 2022 02.
Article in English | MEDLINE | ID: mdl-34624699

ABSTRACT

The mechanical power (MP) is the energy delivered into the respiratory system over time. It can be computed as a direct measurement of the inspiratory area of the airway pressure and volume loop during the respiratory cycle or calculated by "power equations". The absence of a bedside computation limited its widespread use. Recently, it has been developed an automatic monitoring system inside of a mechanical ventilator. PURPOSE: Our aim was to investigate the repeatability and the accuracy of the measured MP at different PEEP values and tidal volume compared with the calculated MP. MATERIAL AND METHODS: MP was measured and calculated in sedated and paralyzed ARDS patients at low and high tidal volume, at 5-10-15 cmH2O of PEEP both in volume and pressure-controlled ventilation. The same measurements were performed twice. RESULTS: Fifty ARDS patients were enrolled. MP was measured and calculated for a total of 300 measurements. The bias and limits of agreement were 0.38 from -1.31 to 2.0 J/min. The measured and calculated MP were similar in each ventilatory condition. CONCLUSIONS: The mechanical power measured by a new automatic real time system implemented in a mechanical ventilator was repeatable and accurate compared with the computed one.


Subject(s)
Respiratory Distress Syndrome , Humans , Lung , Respiration, Artificial , Respiratory Distress Syndrome/therapy , Tidal Volume , Ventilators, Mechanical