Phys Med Biol ; 55(18): 5499-514, 2010 Sep 21.
Article in English | MEDLINE | ID: mdl-20798460

ABSTRACT

Dose measurements following 10 min irradiations with a portable x-ray fluorescence spectrometer composed of a miniature x-ray tube and a silicon PiN diode detector were performed using thermoluminescent dosimeters consisting of LiF:Mg,Ti chips of 3 mm diameter and 0.4 mm thickness. The table-top setup of the spectrometer was used for all measurements. The setup included a stainless steel lid which served as a radiation shield. Two rectangular polyethylene skin/soft tissue phantoms with two cylindrical plaster of Paris bone phantoms were used to study the effect of x-ray beam attenuation and backscatter on the measured dose. Eight different irradiation experiments were performed. The average dose rate values measured with TLD chips within a 1 x 1 cm(2) area were between 4.8 and 12.8 mGy min(-1). The equivalent dose for a 1 x 1 cm(2) skin area was estimated to be 13.2 mSv. The maximum measured dose rate values with a single TLD chip were between 7.5 and 25.1 mGy min(-1). The effective dose corresponding to a proposed arsenic/selenium skin measurement was estimated to be 0.13 microSv for a 2 min irradiation.


Subject(s)
Arsenic/analysis , Radiation Dosage , Selenium/analysis , Skin/chemistry , Skin/radiation effects , Adult , Humans , Male , Phantoms, Imaging , Spectrometry, X-Ray Emission , Thermoluminescent Dosimetry