Opt Lett ; 49(8): 2109-2112, 2024 Apr 15.
Article in English | MEDLINE | ID: mdl-38621088

ABSTRACT

Focused laser differential interferometry (FLDI) is an important diagnostic for measuring density fluctuations in high-speed flows. Currently, however, high dynamic range FLDI is limited to photodiode measurements. In order to spatially resolve multiple locations within complex flows, we present a novel, to the best of our knowledge, refractive-optic imaging FLDI concept that not only produces two-dimensional images without scanning but also reduces the measurement noise floor of those images. To demonstrate this concept, a 33 × 33 grid of FLDI points is first generated using a microlens array. Then, the beams are split and recombined using two polarized Mach-Zehnder interferometers to maximize flexibility in beam separation and optimize signal sensitivity. Next, the FLDI points are collected slightly out of focus on a high-speed camera in order to increase the number of pixels n per FLDI point, thereby reducing noise floor by n. Finally, an under-expanded jet with a characteristic screech at 14.1 kHz is tested with the imaging FLDI setup, showing clear barrel and reflected shock features as well as spatially varying turbulence densities. Overall, this unique concept enables the creation of reduced-noise-floor, two-dimensional FLDI datasets for the study of supersonic and hypersonic flows.