Sensors (Basel) ; 24(9)2024 Apr 25.
Article in English | MEDLINE | ID: mdl-38732830

ABSTRACT

The BC501A sensor is a liquid scintillator frequently used in nuclear physics for detecting fast neutrons. This paper describes a hardware implementation of digital pulse shape analysis (DPSA) for real-time analysis. DPSA is an algorithm that extracts the physically relevant parameters from the detected BC501A signals. The hardware solution is implemented in a MicroTCA system that provides the physical, mechanical, electrical, and cooling support for an AMC board (NAMC-ZYNQ-FMC) with a Xilinx ZYNQ Ultrascale-MP SoC. The Xilinx FPGA programmable logic implements a JESD204B interface to high-speed ADCs. The physical and datalink JESD204B layers are implemented using hardware description language (HDL), while the Xilinx high-level synthesis language (HLS) is used for the transport and application layers. The DPSA algorithm is a JESD204B application layer that includes a FIR filter and a constant fraction discriminator (CFD) function, a baseline calculation function, a peak detection function, and an energy calculation function. This architecture achieves an analysis mean time of less than 100 µs per signal with an FPGA resource utilization of about 50% of its most used resources. This paper presents a high-performance DPSA embedded system that interfaces with a 1 GS/s ADC and performs accurate calculations with relatively low latency.